7. Maintenance
7-4 Performing calibration
118 / 168
7-4 Performing calibration
7-4-1 Calibration preparation
Before performing calibration (zero calibration and span adjustment), prepare the calibration gas.
Also connect the calibration equipment required to the product before starting.
Zero calibration gas
Span adjustment gas (in gas sampling bag or gas can)
Gas sampling bag for exhaust gas
Calibration adapter or splash guard
<When using the calibration adapter>
NOTE
Use a flow rate of 0.5 ± 0.1 L/min when calibrating. Calibration accuracy may be reduced if calibration is
performed using a different flow rate.
If the suction cap and calibration adapter are left fitted for 10 minutes or more in a windless state, gas
may accumulate in the pipes and on the caps, possibly causing readings to increase. If this occurs, use
cleansing air to return the reading to zero. Detection will not be available during this time, as the sensor is
shut off from the atmosphere.
<Using a gas sampling bag>
RI
EC
Product (SD-3)
Flowmeter
Pump
To exhaust
port
Out
P
Calibration
adapter
Needle valve
Gas
sampling
bag
In
Содержание GD-3EC
Страница 28: ...2 Important Safety Information 2 4 Safety information 28 168 Connection examples GD 3RI GD 3ECB SD 3SC SD 3SC...
Страница 169: ...11 Appendix Revision History 169 168...
Страница 170: ...11 Appendix Revision History 170 168...
Страница 171: ...11 Appendix Revision History 171 168...
Страница 172: ...11 Appendix Revision History 172 168...
Страница 173: ...11 Appendix Revision History 173 168...
Страница 174: ...11 Appendix Revision History 174 168...
Страница 175: ...11 Appendix Revision History 175 168...
Страница 176: ...11 Appendix Revision History 176 168...
Страница 177: ...11 Appendix Revision History 177 168...