4. Installation
4-6 Connecting the grounding terminal
56 / 168
4-6 Connecting the grounding terminal
WARNING
Be sure to ground the product before turning on the power.
To ensure stable operation and safety, be sure to ground the product. Never connect grounding wires to
gas pipes.
Make sure the grounding is Class D equivalent (grounding resistance not exceeding 100 Ω).
Use cable lugs on the grounding wire to ensure safe grounding with the cable free of slack and twisting.
Do not ground to the same locations as high frequency or high voltage devices or nearby.
Connect the grounding bolt to the local grounding terminal.
NOTE
The grounding bolt is marked with the
symbol.
<Main unit (SD-3)>
<Main unit (SD-3SC)>
<Remote sensor head (GD-3)>
Grounding bolt (M4)
Grounding bolt (M4)
Grounding bolt (M4)