7. Maintenance
7-3 Performing tests (test mode)
112 / 168
7-3 Performing tests (test mode)
Test mode is used for performing various tests, such as gas tests and alarm tests.
Item No.
Setting item
2-0
Test mode
2-0.0
Gas test
2-0.1
Alarm test
2-0.2
Fault test
2-0.3
LED test
2-0.4
RAM test
7-3-1 Performing gas test
Perform a gas test.
Before starting the gas test, prepare the gas to be used and connect to the product.
For information on gas test preparation, refer to ‘7-4-1 Calibration preparation’.
WARNING
Depending on the settings, contact and gas concentration outputs may be activated during gas testing,
even in maintenance mode. Take care during gas testing.
1 Press the ▲ or ▼ key in maintenance mode several times
to select [2-0], then press the SET key.
2 Select [2-0.0], then press the SET key.
3 Introduce the gas to be used in the gas test.
4 Shut off the gas test gas once the gas concentration
reading has stabilized.
5 Once the reading has returned to [0.0] (or [20.9] (vol%) for
oxygen), press the MENU/ESC key.
Returns to the [2-0.0] display.
Содержание GD-3EC
Страница 28: ...2 Important Safety Information 2 4 Safety information 28 168 Connection examples GD 3RI GD 3ECB SD 3SC SD 3SC...
Страница 169: ...11 Appendix Revision History 169 168...
Страница 170: ...11 Appendix Revision History 170 168...
Страница 171: ...11 Appendix Revision History 171 168...
Страница 172: ...11 Appendix Revision History 172 168...
Страница 173: ...11 Appendix Revision History 173 168...
Страница 174: ...11 Appendix Revision History 174 168...
Страница 175: ...11 Appendix Revision History 175 168...
Страница 176: ...11 Appendix Revision History 176 168...
Страница 177: ...11 Appendix Revision History 177 168...