Analysis
R&S
®
FSV/A3000 I/Q
Analyzer
92
User Manual 1178.8536.02 ─ 01
"Power"
Activates linear power averaging.
The power level values are converted into unit Watt prior to averag-
ing. After the averaging, the data is converted back into its original
unit.
Use this mode to average power values in Volts or Amperes correctly.
In particular, for small VBW values (smaller than the RBW), use
power averaging mode for correct power measurements in FFT
sweep mode.
Remote command:
Predefined Trace Settings - Quick Config
Commonly required trace settings have been predefined and can be applied very
quickly by selecting the appropriate button.
Function
Trace Settings
Preset All Traces
Trace 1:
Clear Write
Traces 2-6:
Blank
Set Trace Mode
Max | Avg | Min
Trace 1:
Max Hold
Trace 2:
Average
Trace 3:
Min Hold
Traces 4-6:
Blank
Set Trace Mode
Max | ClrWrite | Min
Trace 1:
Max Hold
Trace 2:
Clear Write
Trace 3:
Min Hold
Traces 4-6:
Blank
Trace 1 / Trace 2 / Trace 3 / Trace 4 (Softkeys)
Displays the "Traces" settings and focuses the "Mode" list for the selected trace.
Remote command:
DISPlay[:WINDow<n>][:SUBWindow<w>]:TRACe<t>[:STATe]
Copy Trace
Access
: "Overview" > "Analysis" > "Traces" > "Copy Trace"
Copies trace data to another trace.
The first group of buttons (labeled "Trace 1" to "Trace 6" ) selects the source trace. The
second group of buttons (labeled "Copy to Trace 1" to "Copy to Tace 6" ) selects the
destination.
Remote command:
Trace Settings