Remote Commands to Perform Measurements with I/Q Data
R&S
®
FSV/A3000 I/Q
Analyzer
239
User Manual 1178.8536.02 ─ 01
IMAG
Marker search functions are performed on the imaginary trace of
the "I/Q" measurement.
MAGN
Marker search functions are performed on the magnitude of the I
and Q data.
*RST:
REAL
Example:
CALC4:MARK:SEAR IMAG
Manual operation:
See
CALCulate<n>:MARKer<m>:X:SLIMits[:STATe]
<State>
This command turns marker search limits on and off for
all
markers in
all
windows.
If you perform a measurement in the time domain, this command limits the range of the
trace to be analyzed.
Suffix:
<n>
.
irrelevant
<m>
irrelevant
Parameters:
<State>
ON | OFF | 0 | 1
OFF | 0
Switches the function off
ON | 1
Switches the function on
Example:
CALC:MARK:X:SLIM ON
Switches on search limitation.
Manual operation:
See
" Search Limits ( Left / Right )"
See
" Deactivating All Search Limits "
CALCulate<n>:MARKer<m>:X:SLIMits:LEFT
<SearchLimit>
This command defines the left limit of the marker search range for
all
markers in
all
windows.
If you perform a measurement in the time domain, this command limits the range of the
trace to be analyzed.
Suffix:
<n>
.
irrelevant
<m>
irrelevant
I/Q Analysis