Remote Commands to Perform Measurements with I/Q Data
R&S
®
FSV/A3000 I/Q
Analyzer
140
User Manual 1178.8536.02 ─ 01
INPut<ip>:ATTenuation:PROTection:RESet
This command resets the attenuator and reconnects the RF input with the input mixer
for the R&S
FSV/A after an overload condition occurred and the protection mechanism
intervened. The error status bit (bit 3 in the
STAT:QUES:POW
status register) and the
INPUT OVLD
message in the status bar are cleared.
The command works only if the overload condition has been eliminated first.
For details on the protection mechanism see
Suffix:
<ip>
.
1 | 2
irrelevant
Example:
INP:ATT:PROT:RES
INPut<ip>:COUPling
<CouplingType>
This command selects the coupling type of the RF input.
The command is not available for measurements with the optional Digital Baseband
Interface.
Suffix:
<ip>
.
1 | 2
irrelevant
Parameters:
<CouplingType>
AC | DC
AC
AC coupling
DC
DC coupling
*RST:
AC
Example:
INP:COUP DC
Manual operation:
See
INPut<ip>:DPATh
<State>
Enables or disables the use of the direct path for frequencies close to 0
Hz.
Suffix:
<ip>
.
1 | 2
irrelevant
Parameters:
<State>
AUTO | OFF
AUTO | 1
(Default) the direct path is used automatically for frequencies
close to 0
Hz.
Configuring I/Q Analyzer Measurements