Remote Commands to Perform Measurements with I/Q Data
R&S
®
FSV/A3000 I/Q
Analyzer
188
User Manual 1178.8536.02 ─ 01
VIDeo
Video mode is available in the time domain and only in the
Spectrum application.
BBPower
Baseband power (for digital input via the optional Digital Base-
band Interface
PSEN
External power sensor
*RST:
IMMediate
Example:
TRIG:SOUR EXT
Selects the external trigger input as source of the trigger signal
Manual operation:
See
See
See
See
See
See
TRIGger[:SEQuence]:TIME:RINTerval
<Interval>
This command defines the repetition interval for the time trigger.
Parameters:
<Interval>
2.0 ms to 5000
Range:
2 ms to 5000 s
*RST:
1.0 s
Default unit: S
Example:
TRIG:SOUR TIME
Selects the time trigger input for triggering.
TRIG:TIME:RINT 50
The sweep starts every 50 s.
Manual operation:
See
8.4.4.2
Configuring the Trigger Output
The following commands are required to send the trigger signal to one of the variable
"TRIGGER INPUT/OUTPUT" connectors on the R&S
FSV/A.
OUTPut<up>:TRIGger<tp>:DIRection
.............................................................................. 188
.................................................................................... 189
..................................................................................189
OUTPut:TRIGger<tp>:PULSe:IMMediate
..........................................................................190
OUTPut:TRIGger<tp>:PULSe:LENGth
............................................................................. 190
OUTPut<up>:TRIGger<tp>:DIRection
<Direction>
This command selects the trigger direction for trigger ports that serve as an input as
well as an output.
Configuring I/Q Analyzer Measurements