Remote Commands to Perform Measurements with I/Q Data
R&S
®
FSV/A3000 I/Q
Analyzer
183
User Manual 1178.8536.02 ─ 01
[SENSe:]FREQuency:CENTer:STEP:AUTO
<State>
This command couples or decouples the center frequency step size to the span.
In time domain (zero span) measurements, the center frequency is coupled to the
RBW.
Parameters:
<State>
ON | OFF | 0 | 1
*RST:
1
Example:
FREQ:CENT:STEP:AUTO ON
Activates the coupling of the step size to the span.
[SENSe:]FREQuency:OFFSet
<Offset>
This command defines a frequency offset.
If this value is not 0
Hz, the application assumes that the input signal was frequency
shifted outside the application. All results of type "frequency" will be corrected for this
shift numerically by the application.
See also
Parameters:
<Offset>
Range:
-100 GHz to 100 GHz
*RST:
0 Hz
Default unit: HZ
Example:
FREQ:OFFS 1GHZ
Manual operation:
See
8.4.4
Triggering
The following remote commands are required to configure a triggered measurement in
a remote environment. More details are described for manual operation in
*OPC should be used after requesting data. This will hold off any subsequent changes
to the selected trigger source, until after the sweep is completed and the data is
returned.
Configuring the Triggering Conditions
...................................................................183
Configuring the Trigger Output
..............................................................................188
.......................................................................................... 191
8.4.4.1
Configuring the Triggering Conditions
The following commands are required to configure a triggered measurement.
Configuring I/Q Analyzer Measurements