Remote Commands to Perform Measurements with I/Q Data
R&S
®
FSV/A3000 I/Q
Analyzer
171
User Manual 1178.8536.02 ─ 01
OUTPut<up>:UPORt[:VALue]
<Value>
This command sets the control lines of the user ports.
The assignment of the pin numbers to the bits is as follows:
Bit
7
6
5
4
3
2
1
0
Pin
N/A
N/A
5
3
4
7
6
2
Bits 7 and 6 are not assigned to pins and must always be 0.
The user port is written to with the given binary pattern.
Suffix:
<up>
.
irrelevant
Parameters:
<Value>
bit values in hexadecimal format
TTL type voltage levels (max. 5V)
Range:
#B00000000 to #B00111111
Example:
OUTP:UPOR #B00100100
Sets pins 5 and 7 to 5 V.
OUTPut<up>:VIDeo:STATe
<State>
Enables or disables output of the displayed video signal (i.e. the filtered and detected
IF signal) at the "Video" output connector.
Suffix:
<up>
.
1..n
Parameters:
<State>
ON | OFF | 0 | 1
OFF | 0
Switches the function off
ON | 1
Switches the function on
*RST:
0
Example:
OUTP:VID:STAT ON
SYSTem:SPEaker[:STATe]
<State>
This command switches the built-in loudspeaker on or off for demodulated signals.
This setting applies only to the current application.
The command is available in the time domain in Spectrum mode and in Analog
Demodulation mode.
Configuring I/Q Analyzer Measurements