Remote Commands to Perform Measurements with I/Q Data
R&S
®
FSV/A3000 I/Q
Analyzer
264
User Manual 1178.8536.02 ─ 01
Suffix:
<n>
.
irrelevant
<m>
irrelevant
Return values:
<PeakPosition>
Position of the peaks on the y-axis. The unit depends on the
measurement.
Usage:
Query only
MMEMory:STORe<n>:PEAK
<FileName>
This command exports the marker peak list to a file.
Secure User Mode
In secure user mode, settings that are stored on the instrument are stored to volatile
memory, which is restricted to 256
MB. Thus, a "memory limit reached" error can occur
although the hard disk indicates that storage space is still available.
To store data permanently, select an external storage location such as a USB memory
device.
For details, see "Protecting Data Using the Secure User Mode" in the "Data Manage-
ment" section of the R&S
FSV3000/ FSVA3000 base unit user manual.
Suffix:
<n>
.
Parameters:
<FileName>
String containing the path,name and extension of the target file.
Example:
MMEM:STOR:PEAK 'test.dat'
Saves the current marker peak list in the file
test.dat
.
Manual operation:
See
8.8
Retrieving Results
The following commands can be used to retrieve the results of the I/Q Analyzer mea-
surement.
Retrieving Results