Remote Commands to Perform Measurements with I/Q Data
R&S
®
FSV/A3000 I/Q
Analyzer
175
User Manual 1178.8536.02 ─ 01
If you set the attenuation manually, it is no longer coupled to the reference level, but
the reference level is coupled to the attenuation. Thus, if the current reference level is
not compatible with an attenuation that has been set manually, the command also
adjusts the reference level.
Suffix:
<ip>
.
1 | 2
irrelevant
Parameters:
<Attenuation>
Range:
see data sheet
Increment: 5 dB (with optional electr. attenuator: 1 dB)
*RST:
10 dB (AUTO is set to ON)
Default unit: DB
Example:
INP:ATT 30dB
Defines a 30 dB attenuation and decouples the attenuation from
the reference level.
Manual operation:
See
INPut<ip>:ATTenuation:AUTO
<State>
This command couples or decouples the attenuation to the reference level. Thus, when
the reference level is changed, the R&S
FSV/A determines the signal level for optimal
internal data processing and sets the required attenuation accordingly.
This function is not available if the optional Digital Baseband Interface is active.
Suffix:
<ip>
.
1 | 2
irrelevant
Parameters:
<State>
ON | OFF | 0 | 1
*RST:
1
Example:
INP:ATT:AUTO ON
Couples the attenuation to the reference level.
Manual operation:
See
INPut<ip>:ATTenuation:AUTO:MODE
<OptMode>
Selects the priority for signal processing
after
the RF attenuation has been applied.
Suffix:
<ip>
.
1 | 2
irrelevant
Parameters:
<OptMode>
LNOise | LDIStortion
LNOise
Optimized for high sensitivity and low noise levels
Configuring I/Q Analyzer Measurements