Remote Commands to Perform Measurements with I/Q Data
R&S
®
FSV/A3000 I/Q
Analyzer
198
User Manual 1178.8536.02 ─ 01
<NumberSamples>
Number of measurement values to record (including the pretrig-
ger samples).
*RST:
1001
Example:
TRAC:IQ:SET NORM,0,32MHz,EXT,POS,0,2048
Reads 2048 I/Q-values starting at the trigger point.
sample rate = 32 MHz
trigger = External
slope = Positive
TRAC:IQ:SET NORM,0,4 MHz,EXT,POS,1024,512
Reads 512 I/Q-values from 1024 measurement points before the
trigger point.
filter type = NORMAL
sample rate = 4 MHz
trigger = External
slope = Positive
Manual operation:
See
TRACe:IQ:SRATe
<SampleRate>
This command sets the final user sample rate for the acquired I/Q data. Thus, the user
sample rate can be modified without affecting the actual data capturing settings on the
R&S
FSV/A.
Parameters:
<SampleRate>
*RST:
32 MHz
Default unit: HZ
Manual operation:
See
TRACe:IQ:TPISample?
This command queries the time offset between the sample start and the trigger event
(trigger point in sample = TPIS). Since the R&S
FSV/A usually samples with a much
higher sample rate than the specific application actually requires, the trigger point
determined internally is much more precise than the one determined from the (down-
sampled) data in the application. Thus, the TPIS indicates the offset between the sam-
ple start and the actual trigger event.
Configuring I/Q Analyzer Measurements