Basics on I/Q Data Acquisition and Processing
R&S
®
FSV/A3000 I/Q
Analyzer
43
User Manual 1178.8536.02 ─ 01
A Flatop window function is used.
Manual mode
:
The RBW is user-definable.
. Since only window
lengths with integer values can be employed, the
sary, to obtain an integer window length value.
If the record length is larger than the window length, multiple windows are combined;
the FFT length is 4096.
A Flatop window function is used.
Advanced FFT mode
The RBW is determined by the
, depending on the selected
4.8.5
FFT Calculation Methods
FFT calculation can be performed using different methods.
Single
In single mode, one FFT is calculated for the entire record length, that means the win-
dow length is identical to the record length.
If the defined
is larger than the record length, zeros are appended to the
captured data to reach the FFT length.
Figure 4-7: FFT parameters for single FFT calculation
Averaging
In averaging mode, several overlapping FFTs are calculated for each record; the
results are combined to determine the final FFT result for the record.
The number of FFTs to be combined is determined by the
and the
Basics on FFT