Configuration
R&S
®
FSV/A3000 I/Q
Analyzer
85
User Manual 1178.8536.02 ─ 01
Frame Count
Determines how many frames are plotted during a single sweep (as opposed to a con-
tinuous sweep). The maximum number of possible frames depends on the history
depth (see
Remote command:
CALCulate<n>:SPECtrogram:FRAMe:COUNt
Clear Spectrogram
Resets the spectrogram result display and clears the history buffer.
This function is only available if a spectrogram is selected.
Remote command:
CALCulate<n>:SPECtrogram:CLEar[:IMMediate]
5.8
Display Configuration
Access
: "Overview" > "Display Config"
The captured signal can be displayed using various evaluation methods. All evaluation
methods available for the current application are displayed in the evaluation bar in
SmartGrid mode.
For a description of the available evaluation methods see
Up to 6 evaluations can be displayed in the I/Q Analyzer at any time, including several
graphical diagrams, marker tables or peak lists.
The selected evaluation method not only affects the result display in a window, but also
the results of the trace data query in remote control (see
5.9
Adjusting Settings Automatically
Some settings can be adjusted by the R&S
FSV/A automatically according to the cur-
rent measurement settings. In order to do so, a measurement is performed. The dura-
tion of this measurement can be defined automatically or manually.
Adjusting Settings Automatically