Remote Commands to Perform Measurements with I/Q Data
R&S
®
FSV/A3000 I/Q
Analyzer
189
User Manual 1178.8536.02 ─ 01
Suffix:
<up>
.
irrelevant
<tp>
Parameters:
<Direction>
INPut | OUTPut
INPut
Port works as an input.
OUTPut
Port works as an output.
*RST:
INPut
Manual operation:
See
OUTPut<up>:TRIGger<tp>:LEVel
<Level>
This command defines the level of the (TTL compatible) signal generated at the trigger
output.
This command works only if you have selected a user defined output with
.
Suffix:
<up>
.
1..n
<tp>
Selects the trigger port to which the output is sent.
2 = trigger port 2 (front)
3 = trigger port 3 (rear)
2 = Trigger 2 Input / Output
Parameters:
<Level>
HIGH
5 V
LOW
0 V
*RST:
LOW
Example:
OUTP:TRIG2:LEV HIGH
Manual operation:
See
OUTPut<up>:TRIGger<tp>:OTYPe
<OutputType>
This command selects the type of signal generated at the trigger output.
Suffix:
<up>
.
1..n
Configuring I/Q Analyzer Measurements