- 170 -
Immediately send measurement result in current measuring unit
Format:
SUI CR LF
Accessible responses:
SUI_I CR LF
- command understood but not accessible at this
moment
MASS FORMAT - mass value in current measuring unit is returned
immediately
Frame format
1 2 3
4
5
6
7-15
16
17 18 19 20 21
S U I
stability
space sign mass space
unit
CR LF
Example:
S U I CR LF
– command from a computer
S U I ? _ - _ _ _ 5 8 . 2 3 7 _ k g _ CR LF
- command carried out,
immediate response of mass value in current measuring unit
Where: _ - space
Switch on continuous transmission in basic measuring unit
Format:
C1 CR LF
Accessible responses:
C1_I CR LF
- command understood but not accessible at this moment
C1_A CR LF
MASS FRAME
- command understood and in progress
- response is mass value in basic measuring unit
Frame format
1 2
3
4
5
6
7-15
16 17 18 19 20 21
S I space stability
marker space character mass space
unit
CR LF
Switch off continuous transmission in basic measuring unit
Format:
C0 CR LF
Accessible responses:
C0_I CR LF
- command understood but not accessible at this moment
C0_A CR LF
- command understood and in progress