![RADWAG APP 3Y Скачать руководство пользователя страница 163](http://html1.mh-extra.com/html/radwag/app-3y/app-3y_user-manual_3934290163.webp)
- 163 -
Example of a report:
.....................................................
--------- Autotest GLP: Report ---------
Balance type PS 3Y
Balance ID 400010
User Admin
Software rev. L0.0.21 S
Date 2012.01.16
Time 09:17:16
----------------------------------------
Number of measurements 10
Reading unit 0.001/0.01 g
Internal weight mass 1402.094 g
Filter Average
Value release Fast & Reliable
----------------------------------------
Deviation for Max. -0.118 g
Repeatability 0.0088 g
Signature
........................................
PROCEDURE:
On pressing a field with test name, the balance displays a message box.
On this level, the user can carry out the following:
Start the following AUTOTESTU GLP process
Preview results of completed autotests, and export all saved test results as
file with extension *.csv
Delete all saved test results
User can abort the process at optional moment by pressing <X> key in the
testing window.
Results of carried out autotests are displayed as a table (each line features date
and time of autotest procedure and its result).
To preview detailed data of an autotest, press a specified line of the table with
test results.
In order to print a result of a single autotest, enter its details (see above) and
generate a printout by pressing <print> key in the upper bar of the balance’s
display.
Autotest results can be exported by pressing <export> field on the level of all
saved reports in a table. Data is sent to a data storage device connected to
balance’s USB port as a file with extension *.csv.