- 141 -
Control number format:
X / y y / M M / d d / H H / m m
,
T
where:
X - control type which takes the following values:
U – control regulated by the act,
Z – control completed by a user,
yy – year of control end,
MM – month of control end,
dd – day of control end,
HH – hour of control end,
mm – minute of control end,
T – control with determining the average tare value.
Balance user can preview data on each process of average tare determination.
Procedure
•
Enter submenu
< Databases>
in accordance with point 27 of this user
manual,
•
Enter
<
Average tare>
database and select a desired record.
List of parameters defined for a control:
1.
Product
[name of a product which packaging is
checked for average tare determining
process]
2.
Status
[status (result) of a process]
3.
Date
[date of carried out process]
4.
Tare
[determined value of packaging tare]
5.
S
[average standard deviation]
6.
0,25 T1
[the value of a criterion for process
result]
7.
No. of measurements
[Number of completed measurements of
tare value]
8.
User
[name of a user carrying out the process]
27.12.
Pipettes
Database of Pipettes contains records and specification on pipettes to be
calibrated.
Procedure
•
Enter submenu < Databases>, and press <
Pipettes> field
•
Press <
Add > soft key should a new pipette be added,
•
If a pipette already exists in the database of pipettes, press field with its
name to enter its specification which is optionally configurable.
Содержание APP 3Y
Страница 2: ...2 DECEMBER 2012...
Страница 15: ...15 3 BALANCE DESIGN 3 1 Dimensions Ulctramicrobalance UYA 3Y series Balance XA 3Y series...
Страница 16: ...16 Balance XA 3Y F series Balance AS 3Y series...
Страница 17: ...17 Balance PS 3Y series...
Страница 18: ...18 Balance APP 3Y series...