- 103 -
If option <Acquire ambient conditions from THB module> is enabled (set
to <YES>, the software automatically acquires data from the THB module
on temperature, humidity and atmospheric pressure. The value of water
temperature has to be inserted manually, and confirmed by pressing <
Accept> key;
7.
The software returns to the main window, and the grey coloured
workspace displays the following data: tested volume and process status
<In progress>. The task bar displays commands and hits for a user on the
following step - <Weighing sample C1/V1/N1>. Description of texts: C1 –
channel number; V1 – number of volume for a channel; N1 – number of
measurement for tested volume;
8.
Follow displayed commands until completing the calibration process. The
grey coloured workspace displays other data on calibration process, i.e.
average volume, values of errors, specific for carried out process;
9.
On accepting the last measurement, a window is displayed with ambient
conditions parameters. Proceed as on process start and insert all ambient
conditions parameters, and press <
Accept> key;
10.
The software automatically generates a report, that is printed and stored
in the database of reports from pipettes calibration (ambient conditions
and water temperature values given in a report are averages from the
initial and final values of a process);
11.
Complete the process by pressing <
> key visible in the bottom bar.
In case of multichannel pipettes, the balance displays a message on
continuing the process for the following channels. On confirming the
message, the balance carries out calibration / testing process for the
following channel. The settings remain unchanged;
12.
The software enables starting the following procedure for the same
pipette or selecting another pipette from the database.
24.6.
Report from completed calibration processes
On completing each calibration process, a report is generated. The report is
saved in a database <
Reports from pipettes calibration
processes
>. The files are named by their date and time (hour) of
calibration process.
Содержание APP 3Y
Страница 2: ...2 DECEMBER 2012...
Страница 15: ...15 3 BALANCE DESIGN 3 1 Dimensions Ulctramicrobalance UYA 3Y series Balance XA 3Y series...
Страница 16: ...16 Balance XA 3Y F series Balance AS 3Y series...
Страница 17: ...17 Balance PS 3Y series...
Страница 18: ...18 Balance APP 3Y series...