- 171 -
Switch on continuous transmission in current measuring unit
Format:
CU1 CR LF
Accessible responses:
CU1_I CR LF
- command understood but not accessible at this moment
CU1_A CR LF
MASS FRAME
- command understood and in progress
- response is mass value in current measuring unit
Frame format:
1 2 3
4
5
6
7-15
16 17 18 19 20 21
S U I stability
marker space character mass space
unit
CR LF
Switch off continuous transmission in current weighing unit
Format:
CU0 CR LF
Accessible responses:
CU0_I CR LF
- command understood but not accessible at this moment
CU0_A CR LF
- command understood and in progress
Set low checkweighing threshold LO
Format:
DH_XXXXX CR LF
, where
XXXXX
– mass format
Accessible responses:
DH_OK CR LF - command carried out
ES CR LF
- command not recognized (incorrect mass format)
Set high checkweighing threshold HI
Format:
UH_XXXXX CR LF
, where
XXXXX
– mass format
Accessible responses:
UH_OK CR LF
- command carried out
ES CR LF
- command not recognized (incorrect mass format)
Give value of low checkweighing threshold LO
Format:
ODH CR LF
Accessible response:
DH_MASS CR LF
– command carried out
Frame format:
1 2
3
4-12
13
14 15 16
17
18 19
D H
space
mass
space
unit
space
CR LF