- 35 -
•
SUM OF ACTIVE,
totalizing applied tare values for a product or a packaging (resulting
from selecting a product or a packaging from corresponding database)
with a possibility of increasing determined tare value by manual tare
introduced manually using balance’s numeric keyboard. On following
determining of tare value for a product or a packaging, the tare value
introduced manually is deactivated.
•
SUM OF ALL,
totalizing all introduced tare values.
•
AUTOTARE
Means of operation:
Each first stable measurement result is tarred. The NET pictogram
flashes on the display. The operator can determine net mass of weighed
load. On taking off the load from balance’s weighing pan, and as the
indication returns to autozero zone, the software automatically
deactivates Tare value.
−
Automatic footer printout
Accessible options:
MODE -
No
– manual footer printout,
Sum of measurements
– the condition for printing a footer is
exceeding mass value set in parameter <Threshold>
Number of measurements
– the condition for printing a footer
is carrying out a pre-defined number of measurements (a
batch) set in parameter <Threshold>
THRESHOLD – setting the value of threshold determining footer
printout.
For option <Sum of measurements> the value is determined
in measuring unit [g], and for option <Number of
measurements> the value is not measurable, and determined
by number of carried out measurements.
−
Printout mode / printout release
•
Function key PRINT / printout release, (manual operation)
Never – printout deactivated
First stable – the first stable measurement is printed
Each stable – all stable measurements are set for printing
Each
– printout of all measurements (stable and
unstable), in case of a verified balance only stale
measurement results are printable (as in setting
<Each stable>)
•
AUTOMATIC MODE
Never – printout deactivated
First stable
– the first stable measurement result obtained after
placing a load on balance’s weighing pan is recorded,
record of the following measurement result takes
place only on unloading the weighing pan, returning
Содержание APP 3Y