- 142 -
List of parameters defined on a pipette:
1.
Name
2.
Code
3.
Model
4.
Tip
5.
Type of volume
6.
Nominal volume
7.
Minimal volume
8.
Number of channels
9.
Type
10.
Tested volumes
27.13.
Reports from pipette calibration processes
Reports contain data on completed pipette calibration processes. In case of
multichannel pipettes, the reports are generated for each channel separately.
Each report can be previewed, searched by date, exported or printed.
Procedure
•
Enter submenu < Databases>, and press <
Reports from pipette
calibration processes >
•
Press field of a desired report record, unless it is visible on the top of the
list, wind the list forward using navigating keys
•
Report’s name consists of date and time of its creation,
e.g.: 2012.03.12 11:12:15
Caution:
It is possible to use an option for searching a report.
List of parameters in a report from pipette calibration process
:
1.
Pipette [name from the database of pipettes]
2.
Serial no. [serial no. inserted during calibration process]
3.
Channel no.
4.
Start date
5.
End date
6.
User [name of a user carrying out calibration]
7.
Client [client name]
8.
Number of measurements
[no. of measurements for each tested
volume]
9.
Operation with ISO 8655
[information if errors comply with the
standard]
10.
Status [information whether the errors are below permissible
value accepted for the tested volume]
11.
Temperature [average temperature value during calibration
process]
12.
Humidity [average humidity value during calibration process]
13.
Pressure [average pressure value during calibration process]
14.
Water temperature [average water temperature during
calibration process]
15.
Z coefficient [value if the Z coefficient for given temperature]
Содержание APP 3Y
Страница 2: ...2 DECEMBER 2012...
Страница 15: ...15 3 BALANCE DESIGN 3 1 Dimensions Ulctramicrobalance UYA 3Y series Balance XA 3Y series...
Страница 16: ...16 Balance XA 3Y F series Balance AS 3Y series...
Страница 17: ...17 Balance PS 3Y series...