- 132 -
On control completion, the software generates a process summary and the
completed control is automatically saved in the balance’s database.
Caution:
A template and an example of a report from product control is provided in point
26.7 of this user manual.
26.6.
A report from determining average tare value
An example of a report:
Report from average tare: U/22/08/12/14/21/T
----------------------------------------
Balance type PS 3Y
Max capacity 1000 g
Reading unit 0.001 g
Balance ID 9908684
Date 2012.08.22 14:21:10
Product 1
Tare 10.0002 g
0.25 T1 0.225 g
No. of measurements 10
Standard deviation 0.000632 g
Result Positive
Measurements
1. 9.999 g
2. 10.000 g
3. 10.000 g
4. 10.000 g
5. 10.000 g
.
.
.
10. 10.001 g
----------------------------------------
Signature
........................................
Report template:
The balance user in the submenu
<
Printouts/
Printout template of a
report from average tare >
can edit a template of a report from product
control. The variables attributed to <YES> are printed in a report.