Control start:
•
Starting a control requires logging in a user with corresponding access level.
Caution:
3.
Starting a control requires that an access level of the logged user is at
least <user>. If a logged user or an anonymous user have access level
set to <guest> then on control start the software opens a message box
informing on: <Unauthorized access>.
4.
Logging procedure is described in point 9 of this user manual, and
procedure determining access level for each user is provided in point 13 of
this user manual.
•
Save general parameter settings on the
< PGC>
mode operation.
Description in point 26.2 of this user manual.
Select a needed product (by pressing a soft key <
> visible in the bottom bar
or using balance’s keyboard) with correctly determined data on control process:
Caution:
Data on a product indispensable to fill in before starting a control process are
provided in the below table (insert data in accordance with product type and
control type to be carried out, e.g.: density of liquid):
PGC
mode
Available control types: Non-destructive Average Tare,
Non-destructive Empty-Full, Destructive Full-Empty,
Destructive Empty-Full.
Lot
A measuring lot for a control type: Non-destructive
Empty-Full, Destructive Full-Empty, Destructive Empty-
Full
Mass
Nominal product mass
Tare
Packaging mass in an adjustment unit for control type:
Non-destructive Average Tare
Unit
Measuring unit of a product: [g] or [ml]
Density
Product density with respect to product mass in [ml]
(the range of entered values must hold from 0,1g/cm
3
to 5g/cm
3
)