- 85 -
On selecting the formulation mode, the display contains the following
quick access keys in the bottom bar:
1.
Setup – access to balance’s menu
2.
Print header
3.
Print footer
4.
Formulation – selection of a formulation from corresponding database of
formulas
5.
Formulation multiplier (Formulation start if option <Multiplier editing> is
set to <NO>, i.e. disabled)
6.
Target mass
7.
Formulation without a database
22.1.
Additional settings of formulation mode
The additional settings enable adjusting the working mode to user’s needs and
requirements. To access the setting follow below procedure:
Procedure:
1.
Press grey coloured workspace,
2.
The display indicates menu of: Settings, Keys, Information, Printout
and Profile,
3.
Press <Settings> key,
4.
The display indicates functions related to formula making process.
Formula making process is related to the following functions:
−
Automatic ingredients’ naming
:
o
YES
o
NO
−
Apply tare from database of tare values
:
o
YES, each of used ingredients has a tare value assigned in
accordance with a product saved in the database of products;
o
NO, tare is not applied
−
Means of verifying an ingredient
:
The option enables selecting a relation used for determining correct
mass of ingredients during formula making process.
TOLERANCE/TOLERANCE THRESHOLDS
: balance software accepts