- 157 -
Application of outputs:
The set of outputs is used to SIGNAL the status of measurement result.
Change of logic status of the output system, e.g. from [0] to [1] takes place
on meeting a requirement assigned to the output.
Each of 4 outputs is connectable to the following data:
Available options
Balance status, which switches the logic of
output
None
-------
Stable
Each stable measurement switches the logic of
the system
MIN stable
Stable measurement below [MIN] threshold
switches the logic of the system,
MIN unstable
Unstable measurement below [MIN] threshold
switches the logic of the system,
OK stable
Stable measurement within [MIN] [MAX]
thresholds switches the logic of the system,
OK unstable
Unstable measurement within [MIN] [MAX]
thresholds switches the logic of the system,
MAX stable
Stable measurement below [MAX] threshold
switches the logic of the system,
MAX unstable
Unstable measurement below [MAX] threshold
switches the logic of the system
Confirmation of cycle
completion
End of procedure
Procedure
- Press
key and press: “ Inputs / Outputs” key.
- Enter submenu < Inputs / Outputs>, select and press <
Outputs> key
- Enter editing mode of a selected output, which opens list of available
functions,
- Select a function from displayed list and return to weighing with saving
carried out changes.