- 38 -
•
On determining the start mass, the balance display’s a message box:
•
Acting according to the command, place an ordered weight/standard on
balance’s weighing pan and press
key,
•
On completing the adjustment procedure the balance indicates a message
box:
•
On confirming the message box with
key, the balance returns to
weighing mode.
11.3.
User adjustment
User adjustment is carried out with an optional standard, which mass ranges
between 0,15 Max and Max. Adjustment procedure is compatible with the
external adjustment process, but before its start the software opens a
message box for entering mass of a standard used for user adjustment.
CAUTION:
User adjustment is available only in balances which are not subject to conformity
assessment (verification).
In order to start user adjustment, enter submenu < Adjustment >, and select
option: “ User adjustment”. Then follow the commands indicated on
balance’s display.
11.4.
Adjustment test
< Adjustment test > function enables comparing the result of internal automatic
adjustment with the value of internal weight saved in balance’s factory
parameters. The comparison is used for determining balance’s sensitivity drift
over time.