Pioneer 

PX Series

 Balances 

Instruction Manual

 

 

 

 

 

 

 

 

 
 

 
 
 
 

Содержание Pioneer PX125D

Страница 1: ...P Pioneer PX Series Balances Instruction Manual...

Страница 2: ......

Страница 3: ...g the Density of a Solid Using an Auxiliary Liquid other than Water 19 4 5 4 Measuring the Density of a Liquid Using a Calibrated Sinker 20 4 5 5 Measuring the Density of Porous Material Using Oil 21...

Страница 4: ...ory Reset 35 5 9 Lockout 35 6 LEGAL FOR TRADE LFT 36 6 1 Settings 36 6 2 Verification 36 6 3 Securing the Menu 36 6 4 Sealing Access to the Balance Settings 36 7 Printing 37 7 1 Connecting Configuring...

Страница 5: ...For useful information about the product Warning Symbols General Hazard Electrical Shock Hazard 1 4 Safety Precautions CAUTION Read all safety warnings before installing making connections or servici...

Страница 6: ...cluded x Balance x Power adapter Attaching plug x Stainless steel pan x Pan support for 0 1 g 0 01 g 0 00001 g model only x Warrenty card 2 2 Select the Location Avoid heat sources rapid temperature c...

Страница 7: ...ight through serial cable Use the USB port to connect to a computer with a USB 2 0 Type A to Type B cable Interface connections on the rear of the balance USB Used to connect to PC only RS232 Used to...

Страница 8: ...n mode Tare x Performs tare operation Secondary Function Press and Hold Off x Zeroing current value Unit x Changes weighing units Mode x Changes application mode Menu Cal x Enters the main menu Calibr...

Страница 9: ...e load will be displayed as a negative number Press Tare to clear Zero Press Zero to zero the balance Dot matrix Display The relevant data in the specific application mode are shown in the dot matrix...

Страница 10: ...sure Item Settings To view or adjust the current settings Capacity Bar When set to On the capacity bar is displayed in the reference field The capacity will not display when the balance is set to zero...

Страница 11: ...he pan to display the weight Press No or Back to increase or decrease the value as desired 7 Press the Function Mode button so that the weight of the 10 samples is used to establish the average piece...

Страница 12: ...alibrated Use Percent Weighing to display the weight of a test object as a percentage of a pre established reference sample The default or last reference weight is displayed Percent Weighing 1 Press a...

Страница 13: ...Dynamic Weighing cycle Use this application to weigh an unstable load such as a moving animal Dynamic Weighing 1 Press and hold the Function Mode button until Dynamic Weighing appears 2 After confirm...

Страница 14: ...countdown the screen shows the time remaining 6 When the countdown ends the result line is displayed and held 7 After the dynamic object is removed the weight will be automatically set to zero and th...

Страница 15: ...his procedure refer to the density kit however you may use whatever lab apparatus that will suit the requirements for density measurements A built in reference density table for water at temperatures...

Страница 16: ...aker on support as shown NOTE Beaker and thermometer are not supplied as part of the density kit z The density Q is the quotient of the mass m and the volume V Density determinations are performed by...

Страница 17: ...revious procedure After loading the additional mass tare the balance and start the weighing again Wait until the balance has reached stability and note the displayed weight P buoyancy of the solid Imp...

Страница 18: ...in liquid V Volume of the sinker P Buoyancy of the sinker in the liquid P A B In DENSITY SETUP set Mode to Liquid Density and enter sinker volume in cc s After weighing the sinker in air and then wei...

Страница 19: ...aterial Off On x Water Temperature 20 by default x Alcohol Temperature 20 by default x Volume of Calibrated Sinker 10 ml by default x Weight of Porous Material 5 000 g x Oil Density 0 80000 g cm3 Four...

Страница 20: ...n the water temperature value entered 1 Weigh the sample in air using the balance and the density determination kit When the symbol of stability appears press the Function Mode button to confirm the w...

Страница 21: ...e test press the Function Mode button to test a new sample Note Lower the sample down into the liquid until it is fully submerged Holder for non floating solids Holder for floating solids 4 5 3 Measur...

Страница 22: ...suring the Density of a Liquid Using a Calibrated Sinker Press and hold the Function Mode button until Density appears Press Yes to enter the Density Determination Item Settings Sample Type Liquid Vol...

Страница 23: ...ker down into the liquid until it is fully submerged 1 cm below the surface of the liquid When the symbol of stability appears press the Function Mode button to confirm the weight of the calibrated si...

Страница 24: ...1 Weigh the oiled sample in the air with the balance and the density determination kit 2 When the symbol of stability appears press the Function Mode button to confirm the weight of the oiled sample...

Страница 25: ...ure below Before turning the balance over remove the pan and draft shield elements if present to prevent damage Do not place the balance on the pan support cone or load cell pins To use this feature r...

Страница 26: ...d Time User Name Factory Reset Time Format Ounce Troy Balance ID Project Name Time Grain Balance Name Brightness Pennyweight User Name Auto Dim Momme Project Name Capacity Bar Mesghal Application Name...

Страница 27: ...justment is needed place a test mass equal to the span calibration value on the pan and note the difference in divisions between the nominal mass value and the actual balance reading If the difference...

Страница 28: ...s displayed 3 Press Yes to begin the span calibration 4 The calibration masses value will be shown in the screen After the display shows Place weights and 100 000 g place weight s of 100 g on the pan...

Страница 29: ...fication tables in section 9 1 With no load on the scale Linearity Calibration can be performed The balance captures the zero point and then prompts for the next weight Continue to follow the instruct...

Страница 30: ...enu to customize balance functionality Note The factory default settings are shown below in bold 5 3 1 Language 5 3 2 Filter Level Set the amount of signal filtering Low faster stabilization time with...

Страница 31: ...2 2017 Thu 5 3 8 Time Format Set the current time format 24HR 12HR 5 3 9 Time Setup Set the time in the current time format For example if the time format is 24HR the time could be set as 08 00 00 5 3...

Страница 32: ...ff The selected setting is locked x Auto Tare is locked at current setting x Graduations are forced to 1 Division and the menu item is hidden Communication Menu Communication Print Settings Print Outp...

Страница 33: ...Display Unit g Gram dwt Pennyweight kg Kilogram mo Momme t Ton msg Mesghal mg Milligram tl H HK tael ug Microgram tl S SG tael ct Carat tl T TW tael N Newton tcl Tical lb Pound tola Tola oz Ounce baht...

Страница 34: ...ms The LSD is the value by which the displayed weight is incremented or decremented Note The conversion factor is used by the balance to convert grams to the custom weighing unit and is defined by ent...

Страница 35: ...p bit 1 7 NO 2 7 data bits no parity stop bit 2 7 ODD 1 7 data bits odd parity stop bit 1 7 ODD 2 7 data bits odd parity stop bit 2 5 5 3 Handshake Set the flow control method None no handshaking Xon...

Страница 36: ...eader is not printed 5 6 7 Date and Time On the date and the time are printed Off neither the date nor the time is printed 5 6 8 Balance ID On the balance ID is printed Off the balance ID is not print...

Страница 37: ...2 Balance Name Set the balance name Alphanumeric settings up to 16 characters are available 5 7 3 User Name Set the user name Alphanumeric settings up to 16 characters are available The default is bla...

Страница 38: ...he local weights and measures official or authorized service agent must perform the verification procedure 6 3 Securing the Menu A slide switch is used to secure the Lock menu settings When the switch...

Страница 39: ...Properties Settings then ASCII Setup Check boxes as illustrated Send line ends Echo typed characters Wrap lines Use RS232 Interface Commands Section 9 6 1 to control the balance from a PC SPDC Softwa...

Страница 40: ...ated at the immediate left of the most significant digit 4 The Unit field contains the unit of measure abbreviation up to 5 characters right justified 5 The Stability field contains the character if t...

Страница 41: ...gnature ____________ Verified By ____________ Header 1 Header 2 Header 3 Header 4 Header 5 07 19 2017 17 57 19 Balance ID B234567890 Balance Name PX5202 User Name Project Name Parts Counting Quantity...

Страница 42: ...ied By ____________ Header 1 Header 2 Header 3 Header 4 Header 5 07 19 2017 18 00 12 Balance ID B234567890 Balance Name PX5202 User Name Project Name Dynamic Weighing Final weight 49 99 g Gross 50 06...

Страница 43: ...ve balance to suitable location Cannot calibrate Calibration Menu locked Approved Mode set to on Unstable environment Incorrect calibration masses Turn Calibration menu lock off Turn Approved Mode off...

Страница 44: ...ecreasing linearly to 50 relative humidity at 40 C x Operability is assured at ambient temperatures between 5 C and 40 C x Mains supply voltage fluctuations up to 10 of the nominal voltage x Installat...

Страница 45: ...1 Custom unit Applications Basic Weighing Parts Counting Percent Weighing Dynamic Weighing Density Determination Platform Size diameter mm 80 80 80 90 90 90 120 Span Calibration Points g 50 100 40 80...

Страница 46: ...sic Weighing Parts Counting Percent Weighing Dynamic Weighing Density Determination Platform Size diameter mm 120 120 120 120 180 180 180 Span Calibration Points g 100 200 150 300 200 400 250 500 400...

Страница 47: ...s Counting Percent Weighing Dynamic Weighing Density Determination Platform Size diameter mm 180 180 180 180 180 180 180 Span Calibration Points g 1500 3000 2000 4000 2500 5000 1000 2000 2000 4000 300...

Страница 48: ...ighing Density Determination Platform Size diameter mm 90 90 120 120 180 180 180 Span Calibration Points g 50 100 100 200 150 300 250 500 1500 3000 2500 5000 2000 4000 Linearity Calibration Points g 5...

Страница 49: ...5 ACCESSORIES DESCRIPTION PART NUMBER Auxiliary Display PAD7 80251396 Density Kit 80253384 Calibrated Sinker for Liquid Density Determination 83034024 USB Interface Cable 83021085 Security Device 8085...

Страница 50: ...PT Prints Tare weight stored in memory ON Brings out of Standby OFF Goes to Standby C Begin Span Calibration IC Begin Internal Calibration same as trigger from calibration menu AC Abort Calibration At...

Страница 51: ...computer using a Universal Serial Bus USB USB devices are categorized into classes such as disk drives digital cameras printers etc Balances do not have a commonly used class so the Ohaus USB interfac...

Страница 52: ...imit the number of COM port designations e g Ohaus MassTracker allows only COM1 2 3 4 it may be necessary to assign one of these port numbers to the new virtual port Example of Windows XP Hardware Wiz...

Страница 53: ...ents vary by jurisdiction the purchaser should contact their local weights and measures office if they are not familiar with the requirements FCC Note This equipment has been tested and found to compl...

Страница 54: ...been damaged by accident or misuse exposed to radioactive or corrosive materials has foreign material penetrating to the inside of the product or as a result of service or modification by other than...

Страница 55: ......

Страница 56: ...s Corporation 7 Campus Drive Suite 310 Parsippany NJ 07054 USA Tel 1 973 377 9000 Fax 1 973 944 7177 With offices worldwide www ohaus com 30372559 P N 30372559 B 2018 Ohaus Corporation all rights rese...