PX Series Balance
EN-19
4.5.2 Measuring the Density of a floating Solid Using Water
1. Press and hold the
Function / Mode
button until
Density
appears. Press
Yes
to enter the
Density
Determination
.
2. In determining density with the balance, the balance
setup and density determination procedures are
basically the same for a floating solid and a non-
floating solid except for the necessary holder (as
shown in the figure) to be used for density
determination.
3. After completion of the test, press the
Function /
Mode
button to test a new sample.
Note:
Lower the sample down into the liquid until it is fully
submerged.
Holder for non-floating
solids
Holder for floating
solids
4.5.3 Measuring the Density of a Solid Using an Auxiliary Liquid other than Water
1. Press and hold the
Function / Mode
button until
Density
appears. Press
Yes
to enter the
Density Determination
.
Item Settings:
x
Sample type: Solid
x
Auxiliary Liquid: Other
x
Porous Material: Off
Set the density of the auxiliary liquid:
2. The default value of the auxiliary liquid is 1.00000 g/cm
3
3. Press
No
or
Back
to increase or decrease the value in
accordance with the actual density of the auxiliary liquid.