Chapter 2: Analyzer Description
ABL77 Service Manual
Calibration,
Continued
Calibration –
phase 2
•
During the second phase of the calibration, the roller wheel is activated along
with valves 2 and 3. This flushes Cal1 solution through the sensor cassette
measuring chamber (see Figure 2-2).
•
A measurement is made on this solution.
•
The waste pump is also activated to transport residual flush solution to the waste
pouch.
Figure 2-2
Continued on next page
2-8