Chapter 3: Troubleshooting
ABL77 Service Manual
Reference Table,
Continued
Error #
Message
Interpretation
Corrective Action
1311
Calibration required before
analysis.
Maximum allowable time
between calibrations has
been reached.
Perform manual calibration
or allow auto calibration to
complete.
1312
QC Level 1 out of range
At least one parameter fell
outside the acceptable range
for the most recent
BG/Lytes Level 1 QC
measurement
•
Perform a manual
calibration and repeat the
BG/Lytes Level 1
measurement
•
If results remain out of
range inactivate this
parameter or replace the
sensor cassette.
QC Level 2 out of range
At least one parameter fell
outside the acceptable range
for the most recent
BG/Lytes Level 2 QC
measurement
•
Perform a manual
calibration and repeat the
BG/Lytes Level 2
measurement
•
If results remain out of
range inactivate this
parameter or replace the
sensor cassette.
QC Level 3 out of range
At least one parameter fell
outside the acceptable range
for the most recent
BG/Lytes Level 3 QC
measurement
•
Perform a manual
calibration and repeat the
BG/Lytes Level 3
measurement
•
If results remain out of
range inactivate this
parameter or replace the
sensor cassette.
QC Level 4 out of range
At least one parameter fell
outside the acceptable range
for the most recent
BG/Lytes Level 4 QC
measurement
•
Perform a manual
calibration and repeat the
BG/Lytes Level 4
measurement
•
If results remain out of
range inactivate this
parameter or replace the
sensor cassette.
Continued on next page
3-20
Содержание ABL 77 Series
Страница 19: ...Chapter 2 Analyzer Description ABL77 Service Manual This page intentionally left blank 2 4 Rev F ...
Страница 26: ...ABL77 Service Manual Chapter 2 Analyzer Description This page intentionally left blank Rev F 2 11 ...
Страница 40: ...ABL77 Service Manual Chapter 2 Analyzer Description This page intentionally left blank Rev F 2 25 ...
Страница 227: ...Chapter 6 Replacements ABL77 Service Manual This page intentionally left blank 6 32 Rev F ...
Страница 333: ...ABL77 Service Manual Chapter 9 Spare Parts Modules and Externals Continued Module diagrams continued Rev F 9 3 ...
Страница 335: ...ABL77 Service Manual Chapter 9 Spare Parts Upper Module Components Continued Upper module component diagram Rev F 9 5 ...
Страница 344: ...Chapter 9 Spare Parts ABL77 Service Manual This page intentionally left blank 9 14 Rev F ...