Chapter 4: Test and Calibration Procedures
ABL77 Service Manual
Battery Charger,
Continued
Battery charger
detection
Follow the steps below to verify the detection of the external battery charger by the
analyzer. Ensure the analyzer is plugged in correctly and the user has verified the
A/C outlet is active.
Step Action
1.
•
Select
Options
from the
Main Menu
•
Enter a
Password
that will allow access to the
Hardware
screen.
Access to this screen requires a password. If a user password is not
available, the generic MK592 may be used.
NOTE:
This password is for Radiometer use only. Do not share this
password with customers.
•
Select
Hardware
from the
Options
menu.
RESULT:
The following screen will appear.
2.
Verify that the
Pak
reading is above 3000 when the charger is plugged
in to the analyzer and connected to an active AC source.
3.
If the reading is low, replace the battery charger. If the reading
remains low, continue to the next section,
Battery Pack
test.
4-44
Rev.
F
Содержание ABL 77 Series
Страница 19: ...Chapter 2 Analyzer Description ABL77 Service Manual This page intentionally left blank 2 4 Rev F ...
Страница 26: ...ABL77 Service Manual Chapter 2 Analyzer Description This page intentionally left blank Rev F 2 11 ...
Страница 40: ...ABL77 Service Manual Chapter 2 Analyzer Description This page intentionally left blank Rev F 2 25 ...
Страница 227: ...Chapter 6 Replacements ABL77 Service Manual This page intentionally left blank 6 32 Rev F ...
Страница 333: ...ABL77 Service Manual Chapter 9 Spare Parts Modules and Externals Continued Module diagrams continued Rev F 9 3 ...
Страница 335: ...ABL77 Service Manual Chapter 9 Spare Parts Upper Module Components Continued Upper module component diagram Rev F 9 5 ...
Страница 344: ...Chapter 9 Spare Parts ABL77 Service Manual This page intentionally left blank 9 14 Rev F ...