Chapter 8: Maintenance
ABL77 Service Manual
Valve Maintenance
Introduction
This procedure contains instructions for maintaining proper valve performance.
NOTE:
This procedure flushes the internal valves with deionized water. This
periodic procedure is meant to prevent a build-up of salt crystals within the valves.
NOTE:
NOTE:
This procedure is limited to preventive maintenance for the valves.
General maintenance for the waste drain and troubleshooting of valves is
described in Chapter 3 - Troubleshooting.
NOTE:
The Long-term Storage procedure, further in this chapter, is also meant to
prevent salt crystal build-up within the valves and should be performed any time
an analyzer will be turned off for prolonged periods. Long-term Storage must
always be performed prior to shipping an analyzer.
Valve
maintenance
Follow these steps to maintain valve performance.
Step Action
1.
Select
Options
from the
Main Menu
.
2.
Enter the
Password
4M5W6X7.
3.
Press
OK
. The
Options
menu is displayed.
4.
Press
Fluidics
from the
Options
screen.
RESULT:
The following screen appears:
Figure 8-1
5.
From the pop-up window press
Calibration Lines
.
6.
Press
Yes
to continue. The
Calibration Lines
screen is displayed.
Continued on next page
8-2
Rev.
F
Содержание ABL 77 Series
Страница 19: ...Chapter 2 Analyzer Description ABL77 Service Manual This page intentionally left blank 2 4 Rev F ...
Страница 26: ...ABL77 Service Manual Chapter 2 Analyzer Description This page intentionally left blank Rev F 2 11 ...
Страница 40: ...ABL77 Service Manual Chapter 2 Analyzer Description This page intentionally left blank Rev F 2 25 ...
Страница 227: ...Chapter 6 Replacements ABL77 Service Manual This page intentionally left blank 6 32 Rev F ...
Страница 333: ...ABL77 Service Manual Chapter 9 Spare Parts Modules and Externals Continued Module diagrams continued Rev F 9 3 ...
Страница 335: ...ABL77 Service Manual Chapter 9 Spare Parts Upper Module Components Continued Upper module component diagram Rev F 9 5 ...
Страница 344: ...Chapter 9 Spare Parts ABL77 Service Manual This page intentionally left blank 9 14 Rev F ...