ABL77 Service Manual
Chapter 4: Test and Calibration Procedures
Testing the Valve Connections
Introduction
Follow these steps to test the valve connections after replacing the valve board
tubing.
NOTE:
Verify that the manifold is properly installed to prevent other fluidic
problems from arising.
NOTE:
NOTE: Flush water or air from a syringe through the sensor cassette luer
ONLY while the waste pump is running. Flushing without activating the waste
pump can disconnect internal tubing.
Step Action
1.
Apply power to the analyzer.
2.
Select the
Hardware
screen from the
Options
menu. Use password
MK592 when entering the
Options
screen to gain access to the
Hardware
screen.
NOTE:
This password is for Radiometer use only. Do not share this
password with customers.
3.
Insert an index card (or equivalent) into the door slot. The door
indicator in the
Hardware
screen will change from 1 to 0. This will
simulate the presence of a sensor cassette and allow the
Pump Cal1
and
Pump Cal2
buttons to function.
4.
Place a soft cloth or gauze over the manifold ports to collect the water
from the following steps.
NOTE:
Be prepared to collect all fluids that exit the manifold.
NOTE:
Observe the valve board and connections for leaks during
this process.
5.
Fill a syringe, with luer adapter, with DI water.
6.
a) Place the syringe at the sensor cassette fitting luer.
b) Press
Aspirate
and flush the waste lines with de-ionized water.
Fluid should exit the waste port of the manifold.
7.
a) Refill the syringe with DI water and replace it onto the cassette
luer
b) Press
Pump Cal1
c) Flush the Cal1 line. Fluid should exit the Cal1 port of the
manifold.
8.
a) Refill the syringe with DI water and replace it onto the cassette
luer
b) Press
Pump Cal2
c) Flush the Cal2 line. Fluid should exit the Cal2 port of the
manifold.
Continued on next page
Rev. F
4-11
Содержание ABL 77 Series
Страница 19: ...Chapter 2 Analyzer Description ABL77 Service Manual This page intentionally left blank 2 4 Rev F ...
Страница 26: ...ABL77 Service Manual Chapter 2 Analyzer Description This page intentionally left blank Rev F 2 11 ...
Страница 40: ...ABL77 Service Manual Chapter 2 Analyzer Description This page intentionally left blank Rev F 2 25 ...
Страница 227: ...Chapter 6 Replacements ABL77 Service Manual This page intentionally left blank 6 32 Rev F ...
Страница 333: ...ABL77 Service Manual Chapter 9 Spare Parts Modules and Externals Continued Module diagrams continued Rev F 9 3 ...
Страница 335: ...ABL77 Service Manual Chapter 9 Spare Parts Upper Module Components Continued Upper module component diagram Rev F 9 5 ...
Страница 344: ...Chapter 9 Spare Parts ABL77 Service Manual This page intentionally left blank 9 14 Rev F ...