ABL77 Service Manual
Chapter 7: Re-Assembly
Opto-Sensor
Introduction
Follow these steps to re-attach the opto-sensor cable as shown in Figure 7-19.
Step Action
1.
From the back of the module, place the opto sensor into the
rectangular opening in the upper right of the housing.
2.
Fasten the sensor in place with four nylon washer and two Phillips
head screws.
3.
Connect the cable to J6 of the valve board with the green wire in the
top position, as shown below.
Figure 7-19
Part Function
1. Opto-sensor cable connection – J6, green wire on top
2. Opto-sensor with cable
3. Opto-sensor nylon washers and screws
Rev. F
7-31