ABL77 Service Manual
Chapter 4: Test and Calibration Procedures
Battery Pack,
Continued
Charge the
battery pack
(continued)
Step Action
12.
Return to the
Main
Menu and select
Shutdown
.
•
Turn off the analyzer.
•
Allow the battery pack to continue charging for the full 6 hour
cycle.
13.
After a full charge cycle, turn on the analyzer.
14.
At the
Main Menu
disconnect the battery charger from the analyzer.
15.
Verify that the
Main Menu
battery level displays 25.
16.
Proceed to the
Hardware
screen and note the
Bat
reading.
17.
Return to the
Main Menu
.
•
Select
Recall
•
Select
2 Point Cal
•
Select a record
•
Print this record 10 times
18.
Return to the
Hardware
screen. Verify the
Bat
reading has
decremented slightly. If the
Bat
reading does not decrement properly,
replace the battery control board.
Rev. F
4-47
Содержание ABL 77 Series
Страница 19: ...Chapter 2 Analyzer Description ABL77 Service Manual This page intentionally left blank 2 4 Rev F ...
Страница 26: ...ABL77 Service Manual Chapter 2 Analyzer Description This page intentionally left blank Rev F 2 11 ...
Страница 40: ...ABL77 Service Manual Chapter 2 Analyzer Description This page intentionally left blank Rev F 2 25 ...
Страница 227: ...Chapter 6 Replacements ABL77 Service Manual This page intentionally left blank 6 32 Rev F ...
Страница 333: ...ABL77 Service Manual Chapter 9 Spare Parts Modules and Externals Continued Module diagrams continued Rev F 9 3 ...
Страница 335: ...ABL77 Service Manual Chapter 9 Spare Parts Upper Module Components Continued Upper module component diagram Rev F 9 5 ...
Страница 344: ...Chapter 9 Spare Parts ABL77 Service Manual This page intentionally left blank 9 14 Rev F ...