ABL77 Service Manual
Chapter 8: Maintenance
Valve Maintenance,
Continued
Valve
maintenance
(continued)
Step Action
27.
Enter the
Hardware
screen
•
Press
Pump Cal1
to prime the internal fluid lines with Cal1 solution
•
Press
Pump Cal2
to prime the internal fluid lines with Cal2 solution
NOTE:
While Cal1 and Cal2 solution are flushing, snap the sensor
cassette pump tubing to help dislodge all air bubbles as shown in
Figure 8-4.
Figure 8-4
28.
Exit the
Hardware
screen and return to the
Main Menu
.
29.
Perform a manual calibration to ensure proper function prior to sample
analysis.
Rev. F
8-5