ABL77 Service Manual
Chapter 8: Maintenance
Long-Term Storage,
Continued
Long-term
storage
(continued)
Step Action
9.
Fill a syringe (with attached luer adapter) from a Sample Path
Obstruction Tool kit (REF: 905-674) with deionized water or sterile
water. Attach it to the front cassette luer of the analyzer (see
Figure 8-7).
NOTE:
Be prepared to collect all fluids that exit the manifold at the
rear of the analyzer in the cal pack cavity.
Figure 8-7
10.
•
Press
Pump
•
While the waste pump is running, gently introduce the deionized
water into the luer. It will exit the waste port of the manifold.
•
Press
Continue
.
11.
Remove the syringe and pull back on the plunger to fill the syringe
with air. Replace the syringe on the cassette luer.
12.
•
Press
Pump
.
•
While the waste pump is running, gently introduce air into the luer.
This will remove residual fluids from the side waste line.
•
Press
Continue
.
13.
Fill the syringe with deionized water or sterile water. Attach it to the
front cassette luer of the analyzer.
14.
•
Press
Cal 1
.
•
While
Cal1 Open
is displayed on the screen, gently introduce the
deionized water into the luer. It will exit the Cal 1 manifold luer.
•
Press
Continue
.
Continued on next page
Rev. F
8-11
Содержание ABL 77 Series
Страница 19: ...Chapter 2 Analyzer Description ABL77 Service Manual This page intentionally left blank 2 4 Rev F ...
Страница 26: ...ABL77 Service Manual Chapter 2 Analyzer Description This page intentionally left blank Rev F 2 11 ...
Страница 40: ...ABL77 Service Manual Chapter 2 Analyzer Description This page intentionally left blank Rev F 2 25 ...
Страница 227: ...Chapter 6 Replacements ABL77 Service Manual This page intentionally left blank 6 32 Rev F ...
Страница 333: ...ABL77 Service Manual Chapter 9 Spare Parts Modules and Externals Continued Module diagrams continued Rev F 9 3 ...
Страница 335: ...ABL77 Service Manual Chapter 9 Spare Parts Upper Module Components Continued Upper module component diagram Rev F 9 5 ...
Страница 344: ...Chapter 9 Spare Parts ABL77 Service Manual This page intentionally left blank 9 14 Rev F ...