ABL77 Service Manual
Chapter 6: Replacements
Electronics Module
Coin Cell Battery
Introduction
This section describes the process for replacing the coin cell battery. This battery
maintains the CPU settings.
NOTE:
When the coin cell battery is disconnected, the CPU CMOS will lose its settings
and must be re-programmed. See the instructions for re-programming the CPU
CMOS settings in
Chapter 4, Test and Calibration Procedures
.
Coin cell battery
Follow these steps to replace the coin cell battery.
Step Action
1.
Remove the electronics module.
2.
Slide the battery out of the coin cell holder (item
1
in figure 6-18).
3.
Slide a new battery into the battery holder.
4.
Replace the electronics module.
5.
Restart the analyzer.
6.
Check the date and time under
Options
. Adjust as necessary.
NOTE:
When resetting the date, always set the year first otherwise
the CMOS will not record the month or day change.
Figure 6-18
Rev. F
6-29
Содержание ABL 77 Series
Страница 19: ...Chapter 2 Analyzer Description ABL77 Service Manual This page intentionally left blank 2 4 Rev F ...
Страница 26: ...ABL77 Service Manual Chapter 2 Analyzer Description This page intentionally left blank Rev F 2 11 ...
Страница 40: ...ABL77 Service Manual Chapter 2 Analyzer Description This page intentionally left blank Rev F 2 25 ...
Страница 227: ...Chapter 6 Replacements ABL77 Service Manual This page intentionally left blank 6 32 Rev F ...
Страница 333: ...ABL77 Service Manual Chapter 9 Spare Parts Modules and Externals Continued Module diagrams continued Rev F 9 3 ...
Страница 335: ...ABL77 Service Manual Chapter 9 Spare Parts Upper Module Components Continued Upper module component diagram Rev F 9 5 ...
Страница 344: ...Chapter 9 Spare Parts ABL77 Service Manual This page intentionally left blank 9 14 Rev F ...