MPS 3 System – Warnings and Cautions
4.4.2. Cardioplegia Delivery
•
The MPS 3 essential performance is designed to perform within the pressure,
flow rate, and temperature accuracies specified in the operators manual. If you
suspect that the pressure, flow rate, or temperature accuracy is affected by
EMC interference, remove power and relocate the device to another area. It
should be noted that volume accuracy is directly dependent on flow rate
accuracy. If performance continues to be affected discontinue use and contact
Quest Medical Field Service.
•
When venting, the operator should be aware that system pressures greater than
750 mmHg will cause flow to stop.
•
During the venting operation, the arrest and additive pumps are inactive and
delivery stops; however, the delivery ratio is maintained at the setting displayed
and temperature control is active. This may unnecessarily heat, cool or
medicate the volume in the reservoir.
•
DO NOT OPEN
the door while the MPS 3 System is delivering cardioplegia. All
valves will close, flow will stop and the pistons will retract to their home positions
(safe state).
•
The operator is responsible for maintaining safe operating flow rates.
•
The ECG connection on the MPS 3 System is only an activity monitor and is
NOT A MAIN MONITORING DEVICE. Any decisions in care should be made by
a main monitor and not the MPS 3 System display.
•
DO NOT USE
the ports on the MPS 3 Controller during a procedure. This is
only for historical data gathering and other file operations.
10
Содержание MPS 3
Страница 1: ...Quest Medical Inc...
Страница 11: ...MPS 3 System Introduction TTBD Time to be Delivered Term Definition Description 5...
Страница 12: ...MPS 3 System Introduction 6...
Страница 30: ...stem System Specifications and Performance This page intentionally left blank 8...
Страница 42: ...MPS 3 System Warnings and Cautions This page intentionally left blank 12...
Страница 60: ...MPS 3 System System Installation 18...
Страница 64: ...MPS 3 System System Installation Figure 21 Tighten Pole Clamp to Secure Mounting Arm 22...
Страница 90: ...MPS 3 System Startup This page intentionally left blank 9...
Страница 123: ...MPS 3 System Graphical User Interface GUI Overview Figure 80 Auto Mode Screen 33...
Страница 135: ...MPS 3 System Disposable Installation Figure 96 Align the Additive Cartridge 12...
Страница 149: ...MPS 3 System Functional Use Figure 108 New Case Parameters Page 2 Figure 109 New Case Parameters Page 3 3...
Страница 150: ...MPS 3 System Functional Use Figure 110 New Case Parameters Page 4 Figure 111 New Case Parameters Page 5 4...
Страница 151: ...MPS 3 System Functional Use Figure 112 New Case Parameters Page 6 Figure 113 New Case Parameters Page 7 5...
Страница 152: ...MPS 3 System Functional Use Figure 114 New Case Setup Parameters Page 8 6...
Страница 167: ...MPS 3 System Functional Use Figure 130 Dose History Screen 21...
Страница 171: ...MPS 3 System Functional Use Figure 133 Case Report 1 of 4 Figure 134 Case Report Procedure View 25...
Страница 175: ...MPS 3 System Functional Use Figure 136 Case Report 2 of 4 29...
Страница 178: ...MPS 3 System Functional Use 32...
Страница 185: ...MPS 3 System Functional Use Figure 149 Error Report 39...
Страница 187: ...MPS 3 System Functional Use Figure 150 Entering Simulator Mode Warning Figure 151 Simulator Mode 41...
Страница 231: ...MPS 3 System Troubleshooting 40...
Страница 232: ...MPS 3 System Troubleshooting Figure 157 Alarm 314 Figure 158 Alarm 315 41...
Страница 233: ...MPS 3 System Troubleshooting This page intentionally left blank 42...
Страница 234: ...MPS 3 System Troubleshooting 43...
Страница 247: ...MPS 3 System System Maintenance Figure 165 Data Logs Screen 13...
Страница 252: ...MPS 3 System Service 4...
Страница 256: ...MPS 3 System Part Numbers and Accessories 4...
Страница 258: ...MPS 3 System Part Numbers and Accessories 6...
Страница 259: ...MPS 3 System Part Numbers and Accessories 7...
Страница 260: ......
Страница 268: ...MPS 3 System Appendix This page intentionally left blank 8...