MPS 3 System – Troubleshooting
99
Arrest cartridge is empty or absent
Fill Arrest cartridge, press RETEST
DISABLE RETEST
* Choose DISABLE if Arrest
delivery is no longer needed for
procedure
OR
* Refill Arrest Cartridge and
select RETEST
101
Arrest occlusion during prime
Check Arrest line and stopcock
CONFIRM RETEST
* Check arrest delivery line for
clamps, kinks, or occlusions
* Check stopcock position
* If problem persists, replace
arrest delivery line and try again
* If problem still exists after
replacing arrest delivery line,
restart system and try again
* Contact service after multiple
occurrences
102
Arrest cartridge is empty or absent
Arrest pump is Off
REFILL CONFIRM
* Choose CONFIRM if Arrest
delivery is no longer needed for
procedure
OR
* Select REFILL and refill Arrest
Cartridge
103
Arrest occlusion detected
Check Arrest line and stopcock
Arrest pump is Off
CONFIRM RETEST
* Check arrest delivery line for
clamps, kinks, or occlusions
* Check stopcock position
* If problem persists, replace
arrest delivery line and try again
* If problem still exists after
replacing arrest delivery line,
restart system and try again
* Contact service after multiple
occurrences
104
Arrest over-delivery detected
Arrest has been turned off
RETEST CONFIRM
* Select CONFIRM to turn Arrest
pump off and continue
procedure
OR
* Select RETEST to turn Arrest
pump on and continue
procedure
Alarm Code
Message Display
Corrective Action
12
Содержание MPS 3
Страница 1: ...Quest Medical Inc...
Страница 11: ...MPS 3 System Introduction TTBD Time to be Delivered Term Definition Description 5...
Страница 12: ...MPS 3 System Introduction 6...
Страница 30: ...stem System Specifications and Performance This page intentionally left blank 8...
Страница 42: ...MPS 3 System Warnings and Cautions This page intentionally left blank 12...
Страница 60: ...MPS 3 System System Installation 18...
Страница 64: ...MPS 3 System System Installation Figure 21 Tighten Pole Clamp to Secure Mounting Arm 22...
Страница 90: ...MPS 3 System Startup This page intentionally left blank 9...
Страница 123: ...MPS 3 System Graphical User Interface GUI Overview Figure 80 Auto Mode Screen 33...
Страница 135: ...MPS 3 System Disposable Installation Figure 96 Align the Additive Cartridge 12...
Страница 149: ...MPS 3 System Functional Use Figure 108 New Case Parameters Page 2 Figure 109 New Case Parameters Page 3 3...
Страница 150: ...MPS 3 System Functional Use Figure 110 New Case Parameters Page 4 Figure 111 New Case Parameters Page 5 4...
Страница 151: ...MPS 3 System Functional Use Figure 112 New Case Parameters Page 6 Figure 113 New Case Parameters Page 7 5...
Страница 152: ...MPS 3 System Functional Use Figure 114 New Case Setup Parameters Page 8 6...
Страница 167: ...MPS 3 System Functional Use Figure 130 Dose History Screen 21...
Страница 171: ...MPS 3 System Functional Use Figure 133 Case Report 1 of 4 Figure 134 Case Report Procedure View 25...
Страница 175: ...MPS 3 System Functional Use Figure 136 Case Report 2 of 4 29...
Страница 178: ...MPS 3 System Functional Use 32...
Страница 185: ...MPS 3 System Functional Use Figure 149 Error Report 39...
Страница 187: ...MPS 3 System Functional Use Figure 150 Entering Simulator Mode Warning Figure 151 Simulator Mode 41...
Страница 231: ...MPS 3 System Troubleshooting 40...
Страница 232: ...MPS 3 System Troubleshooting Figure 157 Alarm 314 Figure 158 Alarm 315 41...
Страница 233: ...MPS 3 System Troubleshooting This page intentionally left blank 42...
Страница 234: ...MPS 3 System Troubleshooting 43...
Страница 247: ...MPS 3 System System Maintenance Figure 165 Data Logs Screen 13...
Страница 252: ...MPS 3 System Service 4...
Страница 256: ...MPS 3 System Part Numbers and Accessories 4...
Страница 258: ...MPS 3 System Part Numbers and Accessories 6...
Страница 259: ...MPS 3 System Part Numbers and Accessories 7...
Страница 260: ......
Страница 268: ...MPS 3 System Appendix This page intentionally left blank 8...