MPS 3 System – Disposable Installation
8.4.1. Purging the Additive Cartridge Line
Follow these steps to purge excess air from the drug delivery lines:
1.
Select
icon on the Home Screen.
2.
Select the Purge button to initiate the purging process. 1 ml bolus of Additive
solution is delivered.
8.5. Auto Prime for Disabled Arrest and Additive Pumps
The Arrest Agent and Additive pumps must be primed prior to operation to ensure patient
safety and drug delivery accuracy. When the Arrest Agent or Additive pump is disabled, the
K+ or Additive buttons display “Disabled” on the Home Screen. To prime the Arrest Agent or
Additive drug pumps from the Home Screen, follow the steps outlined below.
1. Set the flow rate to zero.
2. Select the Disable button on the corresponding K+ or Additive button on the Home
Screen depending on which pump is disabled. A message window to enable the pump
will be displayed.
Figure 103: Arrest Agent and Additive Disabled
3. Fill the cartridge and then select the ENABLE button on the message window. System
initiates the Auto Priming sequence and Home Screen will be displayed when the
priming process is completed.
The additive pump is automatically turned Off when the
cartridge is empty and the following alarm is displayed:
“Additive cartridge is empty
Additive pump is Off”
When the cartridge is refilled, the pump is automatically
turned on.
One (1) ml of additive is delivered into the heat exchanger
each time the Purge button is selected. Repeatedly using
the Purge button will DELIVER excess additive into the
cardioplegia solution.
Do not use Purge button if air is not present in the additive
cartridge or drug line.
20
Содержание MPS 3
Страница 1: ...Quest Medical Inc...
Страница 11: ...MPS 3 System Introduction TTBD Time to be Delivered Term Definition Description 5...
Страница 12: ...MPS 3 System Introduction 6...
Страница 30: ...stem System Specifications and Performance This page intentionally left blank 8...
Страница 42: ...MPS 3 System Warnings and Cautions This page intentionally left blank 12...
Страница 60: ...MPS 3 System System Installation 18...
Страница 64: ...MPS 3 System System Installation Figure 21 Tighten Pole Clamp to Secure Mounting Arm 22...
Страница 90: ...MPS 3 System Startup This page intentionally left blank 9...
Страница 123: ...MPS 3 System Graphical User Interface GUI Overview Figure 80 Auto Mode Screen 33...
Страница 135: ...MPS 3 System Disposable Installation Figure 96 Align the Additive Cartridge 12...
Страница 149: ...MPS 3 System Functional Use Figure 108 New Case Parameters Page 2 Figure 109 New Case Parameters Page 3 3...
Страница 150: ...MPS 3 System Functional Use Figure 110 New Case Parameters Page 4 Figure 111 New Case Parameters Page 5 4...
Страница 151: ...MPS 3 System Functional Use Figure 112 New Case Parameters Page 6 Figure 113 New Case Parameters Page 7 5...
Страница 152: ...MPS 3 System Functional Use Figure 114 New Case Setup Parameters Page 8 6...
Страница 167: ...MPS 3 System Functional Use Figure 130 Dose History Screen 21...
Страница 171: ...MPS 3 System Functional Use Figure 133 Case Report 1 of 4 Figure 134 Case Report Procedure View 25...
Страница 175: ...MPS 3 System Functional Use Figure 136 Case Report 2 of 4 29...
Страница 178: ...MPS 3 System Functional Use 32...
Страница 185: ...MPS 3 System Functional Use Figure 149 Error Report 39...
Страница 187: ...MPS 3 System Functional Use Figure 150 Entering Simulator Mode Warning Figure 151 Simulator Mode 41...
Страница 231: ...MPS 3 System Troubleshooting 40...
Страница 232: ...MPS 3 System Troubleshooting Figure 157 Alarm 314 Figure 158 Alarm 315 41...
Страница 233: ...MPS 3 System Troubleshooting This page intentionally left blank 42...
Страница 234: ...MPS 3 System Troubleshooting 43...
Страница 247: ...MPS 3 System System Maintenance Figure 165 Data Logs Screen 13...
Страница 252: ...MPS 3 System Service 4...
Страница 256: ...MPS 3 System Part Numbers and Accessories 4...
Страница 258: ...MPS 3 System Part Numbers and Accessories 6...
Страница 259: ...MPS 3 System Part Numbers and Accessories 7...
Страница 260: ......
Страница 268: ...MPS 3 System Appendix This page intentionally left blank 8...