MPS 3 System – System Maintenance
12.1.1.MPS 3 Surface Cleaning and Disinfection
As with all equipment in the operating room, the MPS 3 System external surfaces
should be cleaned and disinfected periodically. Always wipe off spills (blood,
crystalloid, etc.) immediately. Delays may create conditions for microbial growth or
damage the device.
12.1.1.1.Surface Cleaning
To clean the exterior surfaces, follow the steps outlined below.
1. Shutdown the system and unplug the power cord to the MPS 3 Console prior to
cleaning.
2. Open the door to the console.
3. Prepare cleaning solution by adding a 1 tsp (5 ml) of Dish Liquid to 1 gallon of
lukewarm tap water. Mix thoroughly.
4. Wipe the external surfaces of the MPS 3 System with a lint-free cloth
moistened with the prepared soapy water to remove visible soil. Clean all
exposed surfaces of the console exterior and the surfaces inside the door.
5. Use a soft-bristled brush dipped in the prepared soapy water to aid in the
removal of visible soil. Wring any excess solution from the cloth prior to wiping
the console.
6. Wring excess solution from CaviWipes® and gently wipe the external surfaces
of the MPS 3 System until all surfaces are clean from stains, liquid, and dust.
7. Spray CaviCide® on to the external surfaces of the MPS 3 System. Ensure to
pay attention to areas which are difficult to access.
8. Wipe the external surface with CaviWipes® and ensure that the surface
remains visibly wet for 3 minutes.
Avoid OPERATOR INJURY. Ensure that all fingers are
free from pinch / crush hazards while cleaning the MPS
3 System.
Wipe external surfaces thoroughly using only
CaviCide® Wipes or Spray. Do not allow excess solution
to enter housing.
Only use cleaning agents and disinfectants in
compliance with these instructions.
Wear protective gloves when cleaning and disinfecting
the system. Wear protective goggles during the
disinfection process, when disconnecting used tubing
or removing disposables from the system.
Be careful not to blemish the Delivery Temperature
Sensor as this may cause higher than actual Delivery
Temperature displays. Reference Figure 1 for Delivery
Temperature Sensor location.
2
Содержание MPS 3
Страница 1: ...Quest Medical Inc...
Страница 11: ...MPS 3 System Introduction TTBD Time to be Delivered Term Definition Description 5...
Страница 12: ...MPS 3 System Introduction 6...
Страница 30: ...stem System Specifications and Performance This page intentionally left blank 8...
Страница 42: ...MPS 3 System Warnings and Cautions This page intentionally left blank 12...
Страница 60: ...MPS 3 System System Installation 18...
Страница 64: ...MPS 3 System System Installation Figure 21 Tighten Pole Clamp to Secure Mounting Arm 22...
Страница 90: ...MPS 3 System Startup This page intentionally left blank 9...
Страница 123: ...MPS 3 System Graphical User Interface GUI Overview Figure 80 Auto Mode Screen 33...
Страница 135: ...MPS 3 System Disposable Installation Figure 96 Align the Additive Cartridge 12...
Страница 149: ...MPS 3 System Functional Use Figure 108 New Case Parameters Page 2 Figure 109 New Case Parameters Page 3 3...
Страница 150: ...MPS 3 System Functional Use Figure 110 New Case Parameters Page 4 Figure 111 New Case Parameters Page 5 4...
Страница 151: ...MPS 3 System Functional Use Figure 112 New Case Parameters Page 6 Figure 113 New Case Parameters Page 7 5...
Страница 152: ...MPS 3 System Functional Use Figure 114 New Case Setup Parameters Page 8 6...
Страница 167: ...MPS 3 System Functional Use Figure 130 Dose History Screen 21...
Страница 171: ...MPS 3 System Functional Use Figure 133 Case Report 1 of 4 Figure 134 Case Report Procedure View 25...
Страница 175: ...MPS 3 System Functional Use Figure 136 Case Report 2 of 4 29...
Страница 178: ...MPS 3 System Functional Use 32...
Страница 185: ...MPS 3 System Functional Use Figure 149 Error Report 39...
Страница 187: ...MPS 3 System Functional Use Figure 150 Entering Simulator Mode Warning Figure 151 Simulator Mode 41...
Страница 231: ...MPS 3 System Troubleshooting 40...
Страница 232: ...MPS 3 System Troubleshooting Figure 157 Alarm 314 Figure 158 Alarm 315 41...
Страница 233: ...MPS 3 System Troubleshooting This page intentionally left blank 42...
Страница 234: ...MPS 3 System Troubleshooting 43...
Страница 247: ...MPS 3 System System Maintenance Figure 165 Data Logs Screen 13...
Страница 252: ...MPS 3 System Service 4...
Страница 256: ...MPS 3 System Part Numbers and Accessories 4...
Страница 258: ...MPS 3 System Part Numbers and Accessories 6...
Страница 259: ...MPS 3 System Part Numbers and Accessories 7...
Страница 260: ......
Страница 268: ...MPS 3 System Appendix This page intentionally left blank 8...