MPS 3 System – Appendix
device behavior which may be caused by such radiation. The MPS 3 System is intended for use in the
electromagnetic environment specified in the tables below.
17.1.1.Electromagnetic Emission
17.1.2.Electromagnetic Immunity
Guidance and Manufacturer’s Declaration – Electromagnetic Emissions.
The MPS 3 is intended for use in the electromagnetic environment specified below. The customer or the
operator of the device should assure that the device is used in such an environment.
Emissions Test
Compliance
Electromagnetic Environment - Guidance
RF emissions
CISPR 11
Group 1
The MPS 3 System uses RF energy only for its internal
function. Therefore, its RF emissions are very low and
are not likely to cause any interference in nearby
electronic equipment.
RF emissions
CISPR 11
Class B
The product is suitable for use in all establishments,
including domestic establishments and those directly
connected to the public low-voltage power supply network
that supplies buildings used for domestic purpose.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fluctuations/
Flicker emissions
IEC 61000-3-3
Complies
Guidance and Manufacturer’s Declaration – Electromagnetic Immunity
These devices are intended for use in the electromagnetic environment specified below. The customer
or the operator of the device should assure that the device is used in such an environment.
Immunity test
IEC 60601-2, 4th Ed. Test
Level
Compliance level
Electromagnetic
Environment Guidance
Electrostatic
discharge
(ESD)
IEC 61000-4-2
±8 kV contact
±2, ±4, ±8, ±15 kV air
±8 kV contact
±2, ±4, ±8, ±15 kV air
Floor should be wood,
concrete or ceramic tile. If
floors are covered with
synthetic material, the
relative humidity should be
at least 30 %
Electrical
transients /
bursts
IEC 61000-4-4
±0.5, ±1, ±2 kV - 100kHz
repetition frequency
±0.5, ±1, ±2 kV -
100kHz repetition
frequency
Mains power quality should
be that of a typical
commercial and/or hospital
environment
Surge
IEC 61000-4-5
± 1 kV differential mode
± 2 kV common mode voltage
± 1 kV differential
mode
± 2 kV common
mode voltage
The quality of the line power
supply should be that of a
typical commercial or
hospital environment.
2
Содержание MPS 3
Страница 1: ...Quest Medical Inc...
Страница 11: ...MPS 3 System Introduction TTBD Time to be Delivered Term Definition Description 5...
Страница 12: ...MPS 3 System Introduction 6...
Страница 30: ...stem System Specifications and Performance This page intentionally left blank 8...
Страница 42: ...MPS 3 System Warnings and Cautions This page intentionally left blank 12...
Страница 60: ...MPS 3 System System Installation 18...
Страница 64: ...MPS 3 System System Installation Figure 21 Tighten Pole Clamp to Secure Mounting Arm 22...
Страница 90: ...MPS 3 System Startup This page intentionally left blank 9...
Страница 123: ...MPS 3 System Graphical User Interface GUI Overview Figure 80 Auto Mode Screen 33...
Страница 135: ...MPS 3 System Disposable Installation Figure 96 Align the Additive Cartridge 12...
Страница 149: ...MPS 3 System Functional Use Figure 108 New Case Parameters Page 2 Figure 109 New Case Parameters Page 3 3...
Страница 150: ...MPS 3 System Functional Use Figure 110 New Case Parameters Page 4 Figure 111 New Case Parameters Page 5 4...
Страница 151: ...MPS 3 System Functional Use Figure 112 New Case Parameters Page 6 Figure 113 New Case Parameters Page 7 5...
Страница 152: ...MPS 3 System Functional Use Figure 114 New Case Setup Parameters Page 8 6...
Страница 167: ...MPS 3 System Functional Use Figure 130 Dose History Screen 21...
Страница 171: ...MPS 3 System Functional Use Figure 133 Case Report 1 of 4 Figure 134 Case Report Procedure View 25...
Страница 175: ...MPS 3 System Functional Use Figure 136 Case Report 2 of 4 29...
Страница 178: ...MPS 3 System Functional Use 32...
Страница 185: ...MPS 3 System Functional Use Figure 149 Error Report 39...
Страница 187: ...MPS 3 System Functional Use Figure 150 Entering Simulator Mode Warning Figure 151 Simulator Mode 41...
Страница 231: ...MPS 3 System Troubleshooting 40...
Страница 232: ...MPS 3 System Troubleshooting Figure 157 Alarm 314 Figure 158 Alarm 315 41...
Страница 233: ...MPS 3 System Troubleshooting This page intentionally left blank 42...
Страница 234: ...MPS 3 System Troubleshooting 43...
Страница 247: ...MPS 3 System System Maintenance Figure 165 Data Logs Screen 13...
Страница 252: ...MPS 3 System Service 4...
Страница 256: ...MPS 3 System Part Numbers and Accessories 4...
Страница 258: ...MPS 3 System Part Numbers and Accessories 6...
Страница 259: ...MPS 3 System Part Numbers and Accessories 7...
Страница 260: ......
Страница 268: ...MPS 3 System Appendix This page intentionally left blank 8...