MPS 3 System – Troubleshooting
208
System Error
SHUTDOWN RESTART
* Restart System
* Call service after multiple
occurrences
209, 210,
211, 212,
213, 214,
215, 216,
217, 218,
219, 220
Data Logging Error
Consult Service after this case
CONFIRM
* Continue as normal.
* Contact service after multiple
alarm errors
221, 222,
223
Data Log File Transfer
CONFIRM
* Continue as normal.
* Contact service after multiple
alarm errors
224
Arrest pump error
Arrest pump is Off
CONFIRM
* Check Arrest Cartridge to
verify it is properly installed and
fully secured to the Arrest pump
* If the cartridge was secure, try
stopping and restarting
cardioplegia flow
* If error still persists, restart
system and try again
* Contact service after multiple
occurrences
225
Additive pump error
Additive pump is Off
CONFIRM
* Check Additive Cartridge to
verify it is properly installed and
fully secured to the Additive
pump
* If the cartridge was secure, try
stopping and restarting
cardioplegia flow
* If error still persists, restart
system and try again
* Contact service after multiple
occurrences
226
Internal H2O temperature sensor error
Ensure proper Cold water source
Call Service after this Case
CONFIRM
* Continue as normal
OR
* Restart and resume case
* Call service after multiple
occurrences
Alarm Code
Message Display
Corrective Action
27
Содержание MPS 3
Страница 1: ...Quest Medical Inc...
Страница 11: ...MPS 3 System Introduction TTBD Time to be Delivered Term Definition Description 5...
Страница 12: ...MPS 3 System Introduction 6...
Страница 30: ...stem System Specifications and Performance This page intentionally left blank 8...
Страница 42: ...MPS 3 System Warnings and Cautions This page intentionally left blank 12...
Страница 60: ...MPS 3 System System Installation 18...
Страница 64: ...MPS 3 System System Installation Figure 21 Tighten Pole Clamp to Secure Mounting Arm 22...
Страница 90: ...MPS 3 System Startup This page intentionally left blank 9...
Страница 123: ...MPS 3 System Graphical User Interface GUI Overview Figure 80 Auto Mode Screen 33...
Страница 135: ...MPS 3 System Disposable Installation Figure 96 Align the Additive Cartridge 12...
Страница 149: ...MPS 3 System Functional Use Figure 108 New Case Parameters Page 2 Figure 109 New Case Parameters Page 3 3...
Страница 150: ...MPS 3 System Functional Use Figure 110 New Case Parameters Page 4 Figure 111 New Case Parameters Page 5 4...
Страница 151: ...MPS 3 System Functional Use Figure 112 New Case Parameters Page 6 Figure 113 New Case Parameters Page 7 5...
Страница 152: ...MPS 3 System Functional Use Figure 114 New Case Setup Parameters Page 8 6...
Страница 167: ...MPS 3 System Functional Use Figure 130 Dose History Screen 21...
Страница 171: ...MPS 3 System Functional Use Figure 133 Case Report 1 of 4 Figure 134 Case Report Procedure View 25...
Страница 175: ...MPS 3 System Functional Use Figure 136 Case Report 2 of 4 29...
Страница 178: ...MPS 3 System Functional Use 32...
Страница 185: ...MPS 3 System Functional Use Figure 149 Error Report 39...
Страница 187: ...MPS 3 System Functional Use Figure 150 Entering Simulator Mode Warning Figure 151 Simulator Mode 41...
Страница 231: ...MPS 3 System Troubleshooting 40...
Страница 232: ...MPS 3 System Troubleshooting Figure 157 Alarm 314 Figure 158 Alarm 315 41...
Страница 233: ...MPS 3 System Troubleshooting This page intentionally left blank 42...
Страница 234: ...MPS 3 System Troubleshooting 43...
Страница 247: ...MPS 3 System System Maintenance Figure 165 Data Logs Screen 13...
Страница 252: ...MPS 3 System Service 4...
Страница 256: ...MPS 3 System Part Numbers and Accessories 4...
Страница 258: ...MPS 3 System Part Numbers and Accessories 6...
Страница 259: ...MPS 3 System Part Numbers and Accessories 7...
Страница 260: ......
Страница 268: ...MPS 3 System Appendix This page intentionally left blank 8...