MPS 3 System – Graphical User Interface (GUI) Overview
5. The Drain Hex button may be used at the end of the case to drain the water in
the heat exchanger to prevent dripping during removal. H2O Circ has to be
turned off to enable the Drain Hex button.
6. The H2O Prime button may be used to prime the H2O Circulation system. This
is needed to eliminate any air that might be trapped in the circulation system.
7. The Circ flow icon
conveys the health of the water circulation flow. The
green icon is displayed when the water flow is healthy. A flashing gray icon is
displayed if the water flow is suboptimal.
7.4. Function Buttons
The Operator can select the following function buttons from the Home Screen.
•
Vent
•
ECG
•
Hold Volume
•
Graft
•
Menu
•
Protocol
•
Retro
•
Ante
•
Home
•
Auto
•
A dose may also be referred to as a delivery period.
•
A dose starts when the flow is initiated and dose ends when
the flow is stopped for more than programmable ‘off time
delay.’
•
If the flow is resumed before ‘off time delay’ the current dose
resumes.
•
A new dose starts when a delivery route is changed, a new
protocol is launched, Graft mode is activated or deactivated,
a Graft is labeled, or when an ECG event is recorded.
•
A dose always ends when Vent, Recirc, or Hold Vol mode is
activated.
22
Содержание MPS 3
Страница 1: ...Quest Medical Inc...
Страница 11: ...MPS 3 System Introduction TTBD Time to be Delivered Term Definition Description 5...
Страница 12: ...MPS 3 System Introduction 6...
Страница 30: ...stem System Specifications and Performance This page intentionally left blank 8...
Страница 42: ...MPS 3 System Warnings and Cautions This page intentionally left blank 12...
Страница 60: ...MPS 3 System System Installation 18...
Страница 64: ...MPS 3 System System Installation Figure 21 Tighten Pole Clamp to Secure Mounting Arm 22...
Страница 90: ...MPS 3 System Startup This page intentionally left blank 9...
Страница 123: ...MPS 3 System Graphical User Interface GUI Overview Figure 80 Auto Mode Screen 33...
Страница 135: ...MPS 3 System Disposable Installation Figure 96 Align the Additive Cartridge 12...
Страница 149: ...MPS 3 System Functional Use Figure 108 New Case Parameters Page 2 Figure 109 New Case Parameters Page 3 3...
Страница 150: ...MPS 3 System Functional Use Figure 110 New Case Parameters Page 4 Figure 111 New Case Parameters Page 5 4...
Страница 151: ...MPS 3 System Functional Use Figure 112 New Case Parameters Page 6 Figure 113 New Case Parameters Page 7 5...
Страница 152: ...MPS 3 System Functional Use Figure 114 New Case Setup Parameters Page 8 6...
Страница 167: ...MPS 3 System Functional Use Figure 130 Dose History Screen 21...
Страница 171: ...MPS 3 System Functional Use Figure 133 Case Report 1 of 4 Figure 134 Case Report Procedure View 25...
Страница 175: ...MPS 3 System Functional Use Figure 136 Case Report 2 of 4 29...
Страница 178: ...MPS 3 System Functional Use 32...
Страница 185: ...MPS 3 System Functional Use Figure 149 Error Report 39...
Страница 187: ...MPS 3 System Functional Use Figure 150 Entering Simulator Mode Warning Figure 151 Simulator Mode 41...
Страница 231: ...MPS 3 System Troubleshooting 40...
Страница 232: ...MPS 3 System Troubleshooting Figure 157 Alarm 314 Figure 158 Alarm 315 41...
Страница 233: ...MPS 3 System Troubleshooting This page intentionally left blank 42...
Страница 234: ...MPS 3 System Troubleshooting 43...
Страница 247: ...MPS 3 System System Maintenance Figure 165 Data Logs Screen 13...
Страница 252: ...MPS 3 System Service 4...
Страница 256: ...MPS 3 System Part Numbers and Accessories 4...
Страница 258: ...MPS 3 System Part Numbers and Accessories 6...
Страница 259: ...MPS 3 System Part Numbers and Accessories 7...
Страница 260: ......
Страница 268: ...MPS 3 System Appendix This page intentionally left blank 8...