MPS 3 System – Troubleshooting
11.5.Console or Controller Swap Out
It is recommended to change out the MPS 3 Console or MPS 3 Controller only if the MPS 3
System becomes unusable after following the corrective actions and troubleshooting
suggestions in Sections 11.3 and 11.4.
Follow the following instructions if a MPS 3 Console needs to be replaced during a case:
1. Stop Flow.
2. Select Shutdown with Alarm Button or Rear Switch.
3. Clamp Source and Delivery Line.
4. Remove Disposable Set from Console.
5. Unplug Power Cord, Console to Controller Communication Cable from Console,
Circulation Inlet and Outlet Lines, and any additional Accessories connected to the
Console Rear.
6. If on Pole Mount, remove Console Pole Mount locking pin from foot.
7. Replace Console.
8. Connect Power Cord, Console to Controller Communication Cable into Console,
Circulation Inlet and Outlet Lines (ensure adequate water in Circulation System),
and any required Accessories.
9. Turn on Console with Rear Switch and Standby Switch. After System Diagnostics is
completed, select SWAP on Alarm 314.
10. Select SAME CNTLR on Alarm 315.
11. Select RESUME CASE for all previous case parameters to be returned.
12. If using a Hypothermic Reservoir, prime the Circulation System by selecting
Temp Settings
H2O Prime. If using a Heater-Cooler Unit, resume flow from
Heater-Cooler Unit.
Follow the following instructions if a MPS 3 Controller needs to be replaced during a case:
1. Stop Flow.
2. Select Shutdown with Alarm Button or Standby Switch on front of Console (hold for
3 seconds and wait for switch to turn orange).
3. Remove Controller from Mounting Arm.
4. Disconnect Console to Controller Communication Cable from Controller and
Accessory cables.
5. Replace Controller.
6. Connect Console to Controller Communication Cable and required Accessories to
new Controller.
7. Turn on Console with Standby Switch (hold for 3 seconds and wait for switch to turn
blue).
8. After System Diagnostics is completed, select SWAP on Alarm 314.
9. Select DIFF CNTLR on Alarm 315.
10. Select NEW CASE to setup parameters either manually or via a protocol.
39
Содержание MPS 3
Страница 1: ...Quest Medical Inc...
Страница 11: ...MPS 3 System Introduction TTBD Time to be Delivered Term Definition Description 5...
Страница 12: ...MPS 3 System Introduction 6...
Страница 30: ...stem System Specifications and Performance This page intentionally left blank 8...
Страница 42: ...MPS 3 System Warnings and Cautions This page intentionally left blank 12...
Страница 60: ...MPS 3 System System Installation 18...
Страница 64: ...MPS 3 System System Installation Figure 21 Tighten Pole Clamp to Secure Mounting Arm 22...
Страница 90: ...MPS 3 System Startup This page intentionally left blank 9...
Страница 123: ...MPS 3 System Graphical User Interface GUI Overview Figure 80 Auto Mode Screen 33...
Страница 135: ...MPS 3 System Disposable Installation Figure 96 Align the Additive Cartridge 12...
Страница 149: ...MPS 3 System Functional Use Figure 108 New Case Parameters Page 2 Figure 109 New Case Parameters Page 3 3...
Страница 150: ...MPS 3 System Functional Use Figure 110 New Case Parameters Page 4 Figure 111 New Case Parameters Page 5 4...
Страница 151: ...MPS 3 System Functional Use Figure 112 New Case Parameters Page 6 Figure 113 New Case Parameters Page 7 5...
Страница 152: ...MPS 3 System Functional Use Figure 114 New Case Setup Parameters Page 8 6...
Страница 167: ...MPS 3 System Functional Use Figure 130 Dose History Screen 21...
Страница 171: ...MPS 3 System Functional Use Figure 133 Case Report 1 of 4 Figure 134 Case Report Procedure View 25...
Страница 175: ...MPS 3 System Functional Use Figure 136 Case Report 2 of 4 29...
Страница 178: ...MPS 3 System Functional Use 32...
Страница 185: ...MPS 3 System Functional Use Figure 149 Error Report 39...
Страница 187: ...MPS 3 System Functional Use Figure 150 Entering Simulator Mode Warning Figure 151 Simulator Mode 41...
Страница 231: ...MPS 3 System Troubleshooting 40...
Страница 232: ...MPS 3 System Troubleshooting Figure 157 Alarm 314 Figure 158 Alarm 315 41...
Страница 233: ...MPS 3 System Troubleshooting This page intentionally left blank 42...
Страница 234: ...MPS 3 System Troubleshooting 43...
Страница 247: ...MPS 3 System System Maintenance Figure 165 Data Logs Screen 13...
Страница 252: ...MPS 3 System Service 4...
Страница 256: ...MPS 3 System Part Numbers and Accessories 4...
Страница 258: ...MPS 3 System Part Numbers and Accessories 6...
Страница 259: ...MPS 3 System Part Numbers and Accessories 7...
Страница 260: ......