MPS 3 System – System Installation
If the cold water source to be used with the MPS 3 System is a third party Heater
Cooler Unit (HCU), please follow these steps to connect the HCU to the MPS 3
System.
1. Consult the HCU Operation Instructions / User Manual prior to attempting to
connect the HCU to the MPS 3 System the HCU will supply adequate water
flow and water temperatures.
DO NOT
connect a third party HCU to the MPS 3
System if the maximum HCU water pressure exceeds 25psi.
2.
Identify the Inlet and Outlet ports on the HCU cardioplegia water circuit.
Connect
½
” tubing from the Hose Kit to the Inlet and Outlet connectors on the
HCU.
3.
Measure and cut the tubing to the desired length so that it can easily attach to
the MPS 3 Console. Ensure the tubing length is long enough to reach the
console, but not too short to cause kinks or stress the connector or too long to
create a trip hazard.
4. Install hose clamps on the tubing ends. Install one female Hansen fitting
supplied with the Hose Kit into the HCU Outlet – Console Inlet tubing and
secure hose clamp using
¼
” nut driver. Insert the other female Hansen fitting
supplied with the hose kit into the HCU Inlet – Console Outlet tubing and
secure hose clamp using
¼
” nut driver. Flow directional arrows may be applied
to the tubing to help prevent accidental misconnections.
Figure 28: Install Hansen Fittings
29
To H
CU
Inlet
To H
CU O
utlet
Содержание MPS 3
Страница 1: ...Quest Medical Inc...
Страница 11: ...MPS 3 System Introduction TTBD Time to be Delivered Term Definition Description 5...
Страница 12: ...MPS 3 System Introduction 6...
Страница 30: ...stem System Specifications and Performance This page intentionally left blank 8...
Страница 42: ...MPS 3 System Warnings and Cautions This page intentionally left blank 12...
Страница 60: ...MPS 3 System System Installation 18...
Страница 64: ...MPS 3 System System Installation Figure 21 Tighten Pole Clamp to Secure Mounting Arm 22...
Страница 90: ...MPS 3 System Startup This page intentionally left blank 9...
Страница 123: ...MPS 3 System Graphical User Interface GUI Overview Figure 80 Auto Mode Screen 33...
Страница 135: ...MPS 3 System Disposable Installation Figure 96 Align the Additive Cartridge 12...
Страница 149: ...MPS 3 System Functional Use Figure 108 New Case Parameters Page 2 Figure 109 New Case Parameters Page 3 3...
Страница 150: ...MPS 3 System Functional Use Figure 110 New Case Parameters Page 4 Figure 111 New Case Parameters Page 5 4...
Страница 151: ...MPS 3 System Functional Use Figure 112 New Case Parameters Page 6 Figure 113 New Case Parameters Page 7 5...
Страница 152: ...MPS 3 System Functional Use Figure 114 New Case Setup Parameters Page 8 6...
Страница 167: ...MPS 3 System Functional Use Figure 130 Dose History Screen 21...
Страница 171: ...MPS 3 System Functional Use Figure 133 Case Report 1 of 4 Figure 134 Case Report Procedure View 25...
Страница 175: ...MPS 3 System Functional Use Figure 136 Case Report 2 of 4 29...
Страница 178: ...MPS 3 System Functional Use 32...
Страница 185: ...MPS 3 System Functional Use Figure 149 Error Report 39...
Страница 187: ...MPS 3 System Functional Use Figure 150 Entering Simulator Mode Warning Figure 151 Simulator Mode 41...
Страница 231: ...MPS 3 System Troubleshooting 40...
Страница 232: ...MPS 3 System Troubleshooting Figure 157 Alarm 314 Figure 158 Alarm 315 41...
Страница 233: ...MPS 3 System Troubleshooting This page intentionally left blank 42...
Страница 234: ...MPS 3 System Troubleshooting 43...
Страница 247: ...MPS 3 System System Maintenance Figure 165 Data Logs Screen 13...
Страница 252: ...MPS 3 System Service 4...
Страница 256: ...MPS 3 System Part Numbers and Accessories 4...
Страница 258: ...MPS 3 System Part Numbers and Accessories 6...
Страница 259: ...MPS 3 System Part Numbers and Accessories 7...
Страница 260: ......
Страница 268: ...MPS 3 System Appendix This page intentionally left blank 8...