MPS 3 System – Disposable Installation
8.2. Priming and Recirculation
Prior to delivering cardioplegia, the delivery set must be primed to remove all air in the
delivery set and the drug cartridges.
After the disposable is loaded, the MPS 3 System is powered on, and case parameters have
been entered, the MPS 3 System is ready to perform the priming sequence. On completion of
prime the operator has the option to recirculate fluid through the Vent Valve.
When an operator presses the Prime button, the MPS 3 System initiates the priming
sequence. During the automatic priming sequence, the MPS 3 System circulates just enough
cardioplegia solution to clear air bubbles in the delivery line. At the completion of the auto
priming sequence, the manual recirculation mode screen displays.
Prior to starting the priming process, the Arrest Agent and Additive cartridges must be filled. If
a cartridge is not filled a message is displayed giving the operator an option to disable the
specific pump or retest.
To start the Auto Prime sequence, follow the steps outlined below:
1. Select the Prime button on the New Case Setup screen and the start prime screen
displays.
Figure 100: Prime Screen
DO NOT PRIME the delivery set more than 12 hours prior to
surgery. Biological contamination is a risk after more than 12
hours.
When recirculating in Recirc mode or Vent mode, the operator
must ensure that the system has sufficient flow and positive
pressure. If not, the delivery set pouch may drain into the reservoir
which could cause a delay in the next dose of delivery.
15
Содержание MPS 3
Страница 1: ...Quest Medical Inc...
Страница 11: ...MPS 3 System Introduction TTBD Time to be Delivered Term Definition Description 5...
Страница 12: ...MPS 3 System Introduction 6...
Страница 30: ...stem System Specifications and Performance This page intentionally left blank 8...
Страница 42: ...MPS 3 System Warnings and Cautions This page intentionally left blank 12...
Страница 60: ...MPS 3 System System Installation 18...
Страница 64: ...MPS 3 System System Installation Figure 21 Tighten Pole Clamp to Secure Mounting Arm 22...
Страница 90: ...MPS 3 System Startup This page intentionally left blank 9...
Страница 123: ...MPS 3 System Graphical User Interface GUI Overview Figure 80 Auto Mode Screen 33...
Страница 135: ...MPS 3 System Disposable Installation Figure 96 Align the Additive Cartridge 12...
Страница 149: ...MPS 3 System Functional Use Figure 108 New Case Parameters Page 2 Figure 109 New Case Parameters Page 3 3...
Страница 150: ...MPS 3 System Functional Use Figure 110 New Case Parameters Page 4 Figure 111 New Case Parameters Page 5 4...
Страница 151: ...MPS 3 System Functional Use Figure 112 New Case Parameters Page 6 Figure 113 New Case Parameters Page 7 5...
Страница 152: ...MPS 3 System Functional Use Figure 114 New Case Setup Parameters Page 8 6...
Страница 167: ...MPS 3 System Functional Use Figure 130 Dose History Screen 21...
Страница 171: ...MPS 3 System Functional Use Figure 133 Case Report 1 of 4 Figure 134 Case Report Procedure View 25...
Страница 175: ...MPS 3 System Functional Use Figure 136 Case Report 2 of 4 29...
Страница 178: ...MPS 3 System Functional Use 32...
Страница 185: ...MPS 3 System Functional Use Figure 149 Error Report 39...
Страница 187: ...MPS 3 System Functional Use Figure 150 Entering Simulator Mode Warning Figure 151 Simulator Mode 41...
Страница 231: ...MPS 3 System Troubleshooting 40...
Страница 232: ...MPS 3 System Troubleshooting Figure 157 Alarm 314 Figure 158 Alarm 315 41...
Страница 233: ...MPS 3 System Troubleshooting This page intentionally left blank 42...
Страница 234: ...MPS 3 System Troubleshooting 43...
Страница 247: ...MPS 3 System System Maintenance Figure 165 Data Logs Screen 13...
Страница 252: ...MPS 3 System Service 4...
Страница 256: ...MPS 3 System Part Numbers and Accessories 4...
Страница 258: ...MPS 3 System Part Numbers and Accessories 6...
Страница 259: ...MPS 3 System Part Numbers and Accessories 7...
Страница 260: ......
Страница 268: ...MPS 3 System Appendix This page intentionally left blank 8...