Automatic data acquisition
• Connect the VIBXPERT to the multiplexer (see previous section).
• Connect an RPM sensor to the VIBXPERT, if necessary.
• In the route setup (measurement), select the option ‘Autostart
MUX’ (see p 2-16).
• Open the route* with the multiplexer measurement locations.
• Click the first measurement location on the first multiplexer in the
string linen. Note the label of the measurement locations:
• VIBXPERT starts the first measurement task at the first measure-
ment location and saves the result. This procedure continues
automatically until the last measurement task in the string line.
If a measurement task is already completed, the procedure stops
there. You can conduct the respective measurement again or skip
it in order to continue the data acquisition.
Manual data acquisition
• In the route setup (measurement), deactivate the option ‘Autostart
MUX’.
The measurement locations can be processed in any order. Start the
measurements at each measurement location by pressing the button.
The remainder of the procedure is the same as for automatic data
acquisition.
If the connection to the multiplexer is accidentally interrupted dur-
ing the measurement or between measurements, click ‘Reset MUX’
in the menu so that VIBXPERT can again correctly address the mea-
surement locations.
Meas. location
Multiplexer
String line (Multiplexer group)
Location name
in OMNITREND
Left:
Route contains measurement
locations on two multiplexers
Right:
Reset multiplexer
* for machine templates, the procedure
on page 3-13 applies accordingly.
Multiplexer
Note